Skip to the content

FDA Medical Device Recalls (openFDA)

Tearscience, Inc: LipiFlow Thermal Pulsation System/LipiFlow, System Activator (Disposable), Rx Only, Sterile

Recall number
Z-1535-2017
Recalling firm
Tearscience, Inc
Product
LipiFlow Thermal Pulsation System/LipiFlow, System Activator (Disposable), Rx Only, Sterile
Status
Terminated
Initiated
2017-01-30
Posted
2017-03-22
Terminated
2017-11-27
Reason
Labeling: The labeled storage temperature is not consistent with the temperature conditions tested for the entire four year shelf life duration.
Root cause
Labeling Change Control
Action
Tearscience sent an MEDICAL DEVICE LABELING CORRECTION/RECALL letter dated January 30, 2017. .Actions to be Taken by Customer: Within 10 business days of receipt of this letter, please take the following actions: 1) Read the enclosed Summary of Important Labeling Changes for the LipiFlow System. 2) Complete the Acknowledgement and Receipt form at www.TSVerify.pmcprograms.com or return the enclosed completed form by email or mail. 3) If you have stored Activators in an environment that is not temperature controlled (without heat and air conditioning), discontinue use of these Activators and call TearScience Labeling Change Hotline at (844) 872-5658. For any questions regarding this letter, please call (844) 872-5658 Monday through Friday, 8:00AM to 6:00PM, Eastern Time. We want to thank you for being a valuable TearScience partner, and as always would like to thank you for your business, your trust, and your commitment to treating MGD. For further questions please call (919) 459-4831.
Quantity
94,095 units.
Distribution
US (nationwide)
Lot, serial or model codes
Model number LFD-1000 and LFD-1100, All lots
510(k) numbers
["K112704","K133127","K161357"]
PMA numbers
not on file
Product code
ORZ
Firm FEI
3008169506
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register