Skip to the content

FDA Medical Device Recalls (openFDA)

Respironics Novametrix, LLC.: NICO Model 7300 Cardiopulmonary Management System

Recall number
Z-1535-06
Recalling firm
Respironics Novametrix, LLC.
Product
NICO Model 7300 Cardiopulmonary Management System
Status
Terminated
Initiated
2006-08-11
Posted
2006-09-29
Terminated
2012-06-12
Reason
Audible alarm may not sound during an alert condition
Root cause
Other
Action
Respironics notified consignees of the problem with the audible alarm by certified letter on 8/28/06. The correction requires replacement of a speaker upgrade kit.
Quantity
2,432 units
Distribution
Worldwide distribution --- USA and countries of United Arab Emirates, Argentina, Austria, Australia, Belgium, Bulgaria, Bahrain, Brazil, Canada, Canton of Aargau Switzerland, Chile, China, Columbia, Costa Rica, Czech Republic, Germany, Denmark, Algeria, Ecuador, Egypt, Spain, Finland, France, Great Britain, Guam, Hong Kong, Hungary, Ireland, India, Italy, Jamaica, Jordon, Japan, Korea, Kuwait, Lebanon, Lithuania, Morocco, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Peru, Phillippines, Puerto Rico, Portugal, Qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovania, Slovakia, Thailand, Turkey, Taiwan, Uruguay, Venezuala, and South Africa.
Lot, serial or model codes
All serial numbers beginning with 138
510(k) numbers
["K030886","K041450","K982499"]
PMA numbers
not on file
Product code
CCK
Firm FEI
1219324
Firm city
Wallingford
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register