FDA Medical Device Recalls (openFDA)
Beckman Coulter, Inc.: MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: ESL plus, Catalog B1027-101 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
- Recall number
- Z-1534-2015
- Recalling firm
- Beckman Coulter, Inc.
- Product
- MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: ESL plus, Catalog B1027-101 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
- Status
- Terminated
- Initiated
- 2014-08-08
- Posted
- 2015-04-27
- Terminated
- 2015-04-27
- Reason
- Technical Support Bulletin issued in February 2013 did not mention performance limitation for Proteus mirabilis and Imp when using the CLSIM100-S20 revised interpretive criteria.
- Root cause
- Labeling False and Misleading
- Action
- The firm, Siemens, sent an "Urgent Medical Device Correction" letter dated August 7, 2014 to affected customers by trackable method. The letter describe product, problem and actions to be taken. The customers were instructed to confirm receipt of the letter; complete the attached "Field Correction Effectiveness Check" form via fax to Siemens Healthcare Diagnostics at 302-631-8467;TSB 190 has been removed from use and archived from the Siemens Document Library, if you have a hardcopy, discard it appropriately;and follow the listed in instructions and retain and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Customer Care Center; Quality Systems & Compliance Director at 916-374-3031 or email: juntalan@beckman.com or your local Siemens technical support representative.
- Quantity
- Approx ~(216240 boxes U.S. and 84249 boxes OUS) total
- Distribution
- Worldwide Distribution: US (Nationwide) and to countries of: Argentina, Bahamas, Brazil, Canada, China, Ecuador, Egypt, Germany, Lithuania, Mexico, Paraguay, Peru Poland, Portugal, Romania, Russian Federation, Singapore, South Africa, Spain, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates, Uruguay, and Venezuela.
- Lot, serial or model codes
- Catalog number: B1027-101; Siemens Material Number (SMN): 10444755.
- 510(k) numbers
- ["K870074"]
- PMA numbers
- not on file
- Product code
- JWY
- Firm FEI
- 2919016
- Firm city
- West Sacramento
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28