FDA Medical Device Recalls (openFDA)
OrthoPediatrics Corp: 90 Degree Cannulated Infant Blade Plate 25mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractures and osteotomies in all pediatric subgroups (except neonates) and in small stature adults. Specific indications include: intertrochanteric derotation and varus osteotomies, femoral neck and pertrochanteric fractures, intertrochanteric valgus osteotomies, proximal and distal tibial osteotomies and humeral fractures and osteotomies.
- Recall number
- Z-1534-2013
- Recalling firm
- OrthoPediatrics Corp
- Product
- 90 Degree Cannulated Infant Blade Plate 25mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractures and osteotomies in all pediatric subgroups (except neonates) and in small stature adults. Specific indications include: intertrochanteric derotation and varus osteotomies, femoral neck and pertrochanteric fractures, intertrochanteric valgus osteotomies, proximal and distal tibial osteotomies and humeral fractures and osteotomies.
- Status
- Terminated
- Initiated
- 2013-04-19
- Posted
- 2013-06-12
- Terminated
- 2013-09-19
- Reason
- Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.
- Root cause
- Device Design
- Action
- OrthoPediatrics sent an Urgent Voluntary Recall notifiation via telephone and email to all affected consignees. The letter instructed consignees to take possession of any affected devices and quarantine immediately. A Mandatory Reply Form is included with the notification letter to be completed and returned. Please contact Greg Teghtmeyer if you have any questions (574-268-6379).
- Quantity
- 130 devices were shipped
- Distribution
- USA Nationwide Distribution
- Lot, serial or model codes
- All lot numbers are affected. Catalog No. 00-1200-1000
- 510(k) numbers
- ["K110959"]
- PMA numbers
- not on file
- Product code
- HWC
- Firm FEI
- 3006460162
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28