Skip to the content

FDA Medical Device Recalls (openFDA)

Hitachi Medical Systems America Inc: Hitachi Oasis Magnetic Resonance Scanners Software V1.0C The intended use: The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance.

Recall number
Z-1534-2010
Recalling firm
Hitachi Medical Systems America Inc
Product
Hitachi Oasis Magnetic Resonance Scanners Software V1.0C The intended use: The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance.
Status
Terminated
Initiated
2009-09-01
Posted
2010-05-04
Terminated
2012-02-24
Reason
The device has a software error that could potentially lead to a misdiagnosis.
Root cause
Software design (manufacturing process)
Action
Sites have already been corrected. Software updated from V1.0C to V2.0B
Quantity
43
Distribution
Nationwide distribution: AZ, CA, CT, DE, FL, GA, IA, IL, IN, LA, MD, MI, MO, NC, NY, OH, OK, OR, PA, RI, SC, TX, VA, WA, WI, and WY.
Lot, serial or model codes
Model #'s: M001 - M043
510(k) numbers
["K072279"]
PMA numbers
not on file
Product code
IWE
Firm FEI
1528028
Firm city
Twinsburg
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register