ZOLL Circulation, Inc.: Brand Name: AutoPulse NXT Resuscitation System Product Name: AutoPulse NXT Platform Model/Catalog Number: Model 200 (Catalog Number: 8700-001070-1) Software Version: Version 2.0.1 Product Description: The AutoPulse Resuscitation System Model 200 is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. Component: N/A

Recall number
Z-1533-2025
Recalling firm
ZOLL Circulation, Inc.
Product
Brand Name: AutoPulse NXT Resuscitation System Product Name: AutoPulse NXT Platform Model/Catalog Number: Model 200 (Catalog Number: 8700-001070-1) Software Version: Version 2.0.1 Product Description: The AutoPulse Resuscitation System Model 200 is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. Component: N/A
Status
Open, Classified
Initiated
2025-03-04
Posted
2025-04-11
Terminated
not on file
Reason
AutoPulse NXT Resuscitation System may not provide adequate chest compressions or may stop compressions due to code FC1060.
Root cause
Process change control
Action
On March 4, 2025, Zoll Circulation issued a "Medical Device Product Correction Notice" to affected consignees. Zoll asked consignees to take the following actions: 1. Stop using your affected platform(s) immediately. Your patients may not receive high-quality compressions. 2.Complete the Customer Acknowledgment/ No-Charge AP NXT Platform Repair Request Form (page 3)and email it to dlam@zoll.com within 3 days. Please feel free to call me at (408) 419 - 2955 if you have any questions. 3.Return your affected platform(s) to ZOLL as soon as you receive the loaner AP NXT Platform(s) from ZOLL.
Quantity
91
Distribution
US: AL, CA, CT, FL, IL, KS, LA, MN, MO, MT, NC, ND, NM, NY, OH, OK< PA, TX, WI OUS: Canada, Hong Kong,
Lot, serial or model codes
UDI: (01) 00849111003384/Serial number: 1179 1168 1178 1184 1188 1197 1109 1110 1126 1166 1169 1152 1199 1202 1111 1185 1198 1203 1206 1208 1211 1212 1172 1182 1130 1128 1129 1189 1183 1131 1134 1191 1186 1192 1170 1173 1175 1180 1174 1176 1187 1177 1116 1165 1118 1119 1193 1138 1139 1145 1146 1147 1148 1149 1150 1153 1155 1159 1161 1162 1163 1164 1181 1141 1108 1190 1195 1120 1121 1122 1123 1107 1112 1113 1114 1115 1137 1140 1196
510(k) numbers
["K221700"]
PMA numbers
not on file
Product code
DRM
Firm FEI
3003793491
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register