Bard Peripheral Vascular Inc: BARD PERIPHERAL VASCULAR VACCESS PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm. Product Usage: Vaccess PTA Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable stents, self-expanding stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries. Vaccess PTA Balloon Dilatation Catheters are supplied sterile and intended for single use.

Recall number
Z-1533-2013
Recalling firm
Bard Peripheral Vascular Inc
Product
BARD PERIPHERAL VASCULAR VACCESS PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm. Product Usage: Vaccess PTA Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable stents, self-expanding stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries. Vaccess PTA Balloon Dilatation Catheters are supplied sterile and intended for single use.
Status
Terminated
Initiated
2013-05-28
Posted
2013-06-12
Terminated
2013-10-23
Reason
Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have the incorrect balloon diameter labeled on the inflation hub. Specifically, the balloon size printed on the hub may read 7mm x 4cm, when the actual balloon size is 8mm x 4cm. All other product labeling is consistent with the actual 8mm x 4cm balloon size.
Root cause
Mixed-up of materials/components
Action
BARD sent an URGENT MEDICAL DEVICE RECALL NOTIFICATION letters dated May 30, 2013 via FedEx overnight with proof of delivery notice to all affected customers. The letter identified the affected product, reason for recall, actions to be taken, and contact information. Customers were instructed to complete the enclosed Recall and Effectivenss Check Form and fax to the attention of Recall Coordinator in Customer Service at 1-800-994-6772.
Quantity
415
Distribution
US Nationwide Distribution
Lot, serial or model codes
Product Codes and Lot Numbers, respectively:  VA8084R, 93JW0056 VA8084, 937W0090
510(k) numbers
["K073472"]
PMA numbers
not on file
Product code
DQY
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register