Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Company: BD Vacutainer Push Button Blood Collection Set 0.8 x 19 mm x 305 mm 21G x x 12 Catalog 367344

Recall number
Z-1532-2018
Recalling firm
Becton Dickinson & Company
Product
BD Vacutainer Push Button Blood Collection Set 0.8 x 19 mm x 305 mm 21G x x 12 Catalog 367344
Status
Terminated
Initiated
2018-03-20
Posted
not on file
Terminated
2020-04-20
Reason
Products do not meet the labeled sterility claim of a Sterility Assurance Level (SAL) 10-6.
Root cause
Material/Component Contamination
Action
Immediately review your inventory for the specific Catalog (Ref) and lot numbers, and quarantine product subject to the recall. Immediately discontinue the distribution of the affected product. If you have inventory of the recalled product, please complete the attached Business Response Form and return the recalled product following the enclosed packing instruction. This is required so that BD may process your product replacement. If you do not have the recalled product, please complete the attached Business Response Form as well so that BD receives an acknowledgement of your receipt of this recall notification
Quantity
not on file
Distribution
US (Nationwide) Canada, Kuwait, Guyana, Guatemala, Taiwan, UK, Belgium, Sweden, Ireland, Israel, Spain, Switzerland and Lithuania
Lot, serial or model codes
7346627 7348974 7348975 7354740 7354743 7354745 7354764 8005556 8005557 8005814 8010735 8012910 8015923 8015925
510(k) numbers
["K030573"]
PMA numbers
not on file
Product code
JKA
Firm FEI
2243072
Firm city
Franklin Lakes
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28