FDA Medical Device Recalls (openFDA)

Cook Incorporated: Dilator With AQ Hydrophilic Coating - Intended to be used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, biliary and renal systems. REFERENCE PART NUMBER (RPN)/ORDER NUMBER (GPN): 1) JCD10.0-38-20-HC G10207 2) JCD14.0-38-20-HC G10293 3) JCD16.0-38-20-HC G10206 4) JCD6.0-38-20-HC G10285 5) JCD8.0-38-20-HC G10289

Recall number
Z-1531-2024
Recalling firm
Cook Incorporated
Product
Dilator With AQ Hydrophilic Coating - Intended to be used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, biliary and renal systems. REFERENCE PART NUMBER (RPN)/ORDER NUMBER (GPN): 1) JCD10.0-38-20-HC G10207 2) JCD14.0-38-20-HC G10293 3) JCD16.0-38-20-HC G10206 4) JCD6.0-38-20-HC G10285 5) JCD8.0-38-20-HC G10289
Status
Open, Classified
Initiated
2024-03-01
Posted
2024-04-12
Terminated
not on file
Reason
Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
Root cause
Under Investigation by firm
Action
Cook Medical issued Urgent Medical Device Recall letter on 3/1/24 to Risk Management/Recall Administration. . Letter states reason for recall, health risk and action to take: Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is removing potentially affected products from the market. Additionally, a supplier investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue.
Quantity
1,493 units
Distribution
Worldwide distribution - US Nationwide and the countries of AT, AU, BR, CA, CH, DE, ES, GB, HK, IE, IL, IN, IT, NZ, SG.
Lot, serial or model codes
Ref. Number/UDI-DI/Lot Numbers: 1) JCD10.0-38-20-HC (01)00827002102071(17) 15794369 2) JCD14.0-38-20-HC (01)00827002102934(17) 15778297, 15787264 3) JCD16.0-38-20-HC (01)00827002102064(17) 15794375 4)JCD6.0-38-20-HC (01)00827002102859(17) NS15808469 5) JCD8.0-38-20-HC (01)00827002102897(17) 15761259
510(k) numbers
["K183036"]
PMA numbers
not on file
Product code
FGE
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register