Skip to the content

FDA Medical Device Recalls (openFDA)

Menarini Silicon Biosystems: CELLSEARCH CIRCULATING Tumor Cell Kits (IVD) (Product Code 7900001) CELLSEARCH CIRCULATING Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of the epithelial origin (CD45-, EpCAM+, cytokeratins 8, 18+, and/or 19+) in whole blood.

Recall number
Z-1530-2018
Recalling firm
Menarini Silicon Biosystems
Product
CELLSEARCH CIRCULATING Tumor Cell Kits (IVD) (Product Code 7900001) CELLSEARCH CIRCULATING Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of the epithelial origin (CD45-, EpCAM+, cytokeratins 8, 18+, and/or 19+) in whole blood.
Status
Terminated
Initiated
2018-04-04
Posted
not on file
Terminated
2019-04-09
Reason
Unusually high number of total images/unassigned events may require additional time and effort for the operator to review all the images and may cause a delay in reporting the results.
Root cause
Process design
Action
Menarini silicon biosystems sent an Urgent Medical Device Correction Notification dated April 3, 2018. Required Actions. " Please dispose of any affected product in your current inventory . " Report any previous occurrence of unusually high number of total images I unassigned events in patient samples to Customer Technical Support. " Complete and return the enclosed Confirmation of Receipt form no later than April 20,2018. For further question, please call (215) 830-0777.
Quantity
not on file
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : Canada and Italy
Lot, serial or model codes
UDI (01)10888385000012(17)180810(10)S038 (01)10888385000012(17)180810(10) (01)10888385000012(17)180810(10)S038S (01)10888385000012(17)180810(10)S038T (01)10888385000012(17)180810(10)S038U (01)10888385000012(17)180810(10)S038 V  Serial, Batch, Lot # 2038 S038R S038S S038T S038U S038V S038W (0 distributed to the field)
510(k) numbers
["K103502"]
PMA numbers
not on file
Product code
NQI
Firm FEI
3004153557
Firm city
Huntingdon Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register