Menarini Silicon Biosystems: CELLSEARCH CIRCULATING Tumor Cell Kits (IVD) (Product Code 7900001) CELLSEARCH CIRCULATING Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of the epithelial origin (CD45-, EpCAM+, cytokeratins 8, 18+, and/or 19+) in whole blood.
- Recall number
- Z-1530-2018
- Recalling firm
- Menarini Silicon Biosystems
- Product
- CELLSEARCH CIRCULATING Tumor Cell Kits (IVD) (Product Code 7900001) CELLSEARCH CIRCULATING Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of the epithelial origin (CD45-, EpCAM+, cytokeratins 8, 18+, and/or 19+) in whole blood.
- Status
- Terminated
- Initiated
- 2018-04-04
- Posted
- not on file
- Terminated
- 2019-04-09
- Reason
- Unusually high number of total images/unassigned events may require additional time and effort for the operator to review all the images and may cause a delay in reporting the results.
- Root cause
- Process design
- Action
- Menarini silicon biosystems sent an Urgent Medical Device Correction Notification dated April 3, 2018. Required Actions. " Please dispose of any affected product in your current inventory . " Report any previous occurrence of unusually high number of total images I unassigned events in patient samples to Customer Technical Support. " Complete and return the enclosed Confirmation of Receipt form no later than April 20,2018. For further question, please call (215) 830-0777.
- Quantity
- not on file
- Distribution
- Worldwide Distribution - USA (nationwide) and to the countries of : Canada and Italy
- Lot, serial or model codes
- UDI (01)10888385000012(17)180810(10)S038 (01)10888385000012(17)180810(10) (01)10888385000012(17)180810(10)S038S (01)10888385000012(17)180810(10)S038T (01)10888385000012(17)180810(10)S038U (01)10888385000012(17)180810(10)S038 V Serial, Batch, Lot # 2038 S038R S038S S038T S038U S038V S038W (0 distributed to the field)
- 510(k) numbers
- ["K103502"]
- PMA numbers
- not on file
- Product code
- NQI
- Firm FEI
- 3004153557
- Firm city
- Huntingdon Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28