Skip to the content

FDA Medical Device Recalls (openFDA)

Corin USA Limited: Trinity Acetabular Shell Product Usage Total Hip Replacement.

Recall number
Z-1528-2016
Recalling firm
Corin USA Limited
Product
Trinity Acetabular Shell Product Usage Total Hip Replacement.
Status
Terminated
Initiated
2016-01-29
Posted
not on file
Terminated
2019-12-16
Reason
Potential for abrasion of the sterile barrier pouches of the new packaging system.
Root cause
Packaging process control
Action
Corin sent an Urgent Medical Device Recall letter dated February 3, 2016. The letter identified the affected product, problem and actions to be taken. On January 29, 2016, Eight (8) of 11 distributors were spoken to during a Corin National Sales Meeting informing them of the recall. The 3 remaining distributors were notified by phone calls same date. Also, on February 2, 2016 Corin sent a Recall Letter to their Foreign consignee informating them of the recall of the Trinity Acetabular Shell and the actions to be take by the costumers including the return of the Customer Response Form and Contact information as Corin LTD Compliance Director at +44 (0) 1285 649190 Monday thru Friday 8:30am to 5:00pm UK Time
Quantity
629 devices in total
Distribution
Worldwide Distribution - US Nationwide in the states of CA, FL, ID, TX, OK, OH, MA, and NC and the countries of Australia Belgium, Egypt, Germany, Ghana, Italy, Japan, Poland, South Africa and UK.
Lot, serial or model codes
No Hole Shell 048mm Taper Size 2, Lot Codes: 322986, 323743, 320026, 322977, 323744; No Hole Shell 052mm Taper Size 3, Lot Codes: 322396, 322971, 322403, 322984, 322985, 323750.
510(k) numbers
["K123705"]
PMA numbers
not on file
Product code
LZO
Firm FEI
1000131827
Firm city
Tampa
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register