FDA Medical Device Recalls (openFDA)
INO Therapeutics (dba Ikaria): INOmax DSIR, model 10007, Nitric Oxide Delivery System, with Software version 3.0.0. The INOmax DS delivery system delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
- Recall number
- Z-1528-2014
- Recalling firm
- INO Therapeutics (dba Ikaria)
- Product
- INOmax DSIR, model 10007, Nitric Oxide Delivery System, with Software version 3.0.0. The INOmax DS delivery system delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
- Status
- Terminated
- Initiated
- 2014-03-27
- Posted
- 2014-05-02
- Terminated
- 2014-11-21
- Reason
- An anomaly impacts the monitoring system of INOmax DSIR Plus (software 3.0.0). When the integrated pneumatic backup switch is activated with low or no gas flow through the injector module, an intended NO cell failure alarm will occur. In addition, an unintended NO2 cell failure alarm may also occur. The NO/NO2 cell failure alarms are caused by an average NO/NO2 circuit concentration greater tha
- Root cause
- Software Design Change
- Action
- Ikaria sent an Urgent Field Corrective Action letter via FedEx on March 27, 2014. The letter identified the issue, and requested that all INOmax DSIR with software version 3.0.0 be returned to Ikaria. Customers with questions were instructed to contact Ikaria Customer Care at 1-877-566-9466. For questions regarding this recall call 877-566-9466.
- Quantity
- 22
- Distribution
- Nationwide Distribution including DC and VA.
- Lot, serial or model codes
- Serial Numbers: DS20080152, DS20090218, DS20090243, DS20090271, DS20090523, DS20090598, DS20090821, DS20090925, DS20101022, DS20101076, DS20101147, DS20101374, DS20101413, DS20101420, DS20101499, DS20101637, DS20111199, DS20120099, DS20120150, DS20120169, DS20120288, DS20120306.
- 510(k) numbers
- ["K131686"]
- PMA numbers
- not on file
- Product code
- MRN
- Firm FEI
- 3004531588
- Firm city
- Madison
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28