Skip to the content

FDA Medical Device Recalls (openFDA)

Ge Healthcare, Llc: GE, 1009-9002-000 Avance Anesthesia Machine and Monitor

Recall number
Z-1527-2010
Recalling firm
Ge Healthcare, Llc
Product
GE, 1009-9002-000 Avance Anesthesia Machine and Monitor
Status
Terminated
Initiated
2010-03-12
Posted
2010-05-11
Terminated
2012-01-30
Reason
Unanticipated shut-down: A specific lot of GE Healthcare Aisys and Avance machines control board wiring harness have a defect, which can cause the machine to unexpectedly shut down, terminating ventilation, anesthetic delivery, and potentially patient monitoring.
Root cause
Process design
Action
Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated March 12, 2010. The letter was addressed to Chief of Anesthesia, Health Care Administrator/Risk Manager, Director of Biomedical/Clinical Engineering. The letter listed the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Quantity
18 (4 USA, 14 OUS)
Distribution
Worldwide distribution: USA (CA, FL, TN), CANADA, DENMARK, FINLAND, FRANCE, ITALY, SAUDI ARABIA, SOUTH AFRICA, SPAIN, UNITED ARAB EMIRATES, and UNITED KINGDOM.
Lot, serial or model codes
Serial Numbers: ANBN01180, ANBN01187, ANBN01200, ANBN01201, ANBN01211, ANBN01212, ANBN01213, ANBN01220, ANBN01223, ANBN01224, ANBN01250, ANBN01266, ANBN01284, ANBN01285, ANBN01286, ANBN01287, ANBN01288, ANBN01289
510(k) numbers
["K081844"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register