FDA Medical Device Recalls (openFDA)
XTANT MEDICAL INC: OsteoSelect Demineralized Bone Matrix (DBM) Putty 10.0cc in a syringe; STERILE R, Product Code: 359100, Dimensions: ONE(1) PACKAGE - CONTAINING 10.0cc. UDI: 00849777003681 - Product Usage: OsteoSelect DBM Putty is indicated for use as a bone void filler and bone graft substitute for voids or gaps that are not intrinsic to the stability of the bony structure. OsteoSelect DBM Putty is indicated for treatment of surgically created osseous defects or osseous defects from traumatic injury to the bone. OsteoSelect DBM Putty can be used as follows: - Extremities - Posterolateral spine - Pelvis.
- Recall number
- Z-1526-2020
- Recalling firm
- XTANT MEDICAL INC
- Product
- OsteoSelect Demineralized Bone Matrix (DBM) Putty 10.0cc in a syringe; STERILE R, Product Code: 359100, Dimensions: ONE(1) PACKAGE - CONTAINING 10.0cc. UDI: 00849777003681 - Product Usage: OsteoSelect DBM Putty is indicated for use as a bone void filler and bone graft substitute for voids or gaps that are not intrinsic to the stability of the bony structure. OsteoSelect DBM Putty is indicated for treatment of surgically created osseous defects or osseous defects from traumatic injury to the bone. OsteoSelect DBM Putty can be used as follows: - Extremities - Posterolateral spine - Pelvis.
- Status
- Terminated
- Initiated
- 2019-05-03
- Posted
- not on file
- Terminated
- 2020-12-17
- Reason
- The firm received one customer complaint concerning three OsteoSelect Demineralized Bone Matrix (DBM) devices which contained 5.0cc of putty, instead of 10.00cc of putty as labeled.
- Root cause
- Labeling Change Control
- Action
- On May 3, 2019 a customer letter (in an envelope marked URGENT) was Federal Expressed overnight-signature required to the 16 customers who received the 26 devices. The specific actions to be taken by the customer/user that were shipped the product affected by the recall are requested to do the following: 1. Immediately examine their inventory and quarantine any affected products subject to this recall. 2. Complete an enclosed Acknowledgement Form, even if they have no remaining devices in their inventory, and return the completed Acknowledgment Form via mail: Mail: DBM Recall 664 Cruiser Lane Belgrade, MT 59714 Email: mdolan@xtantmedical.com Fax: 1-406-388-3380 3. Contact via email: mdolan@xtantmedical.com, or 1-406-388-0480 ext. 1127, between 8 a.m. and 5 p.m. (Mountain Time) to obtain instructions for returning affected product, and to obtain replacement devices which will be provided at no cost. 4. Package the affected products to be returned. The firm will provide a shipping container (if needed) with a return shipping label. Securely package the affected product(s). Send the container to: ATTN: OSTEOSELECT DBM RECALL (RMA Number) 732 Cruiser Lane Belgrade, MT 59714 For questions, please contact the Director of RA/QA at 1-406-388-0480 ext. 1128, from 8 a.m. to 5 p.m. (Mountain Time).
- Quantity
- 26
- Distribution
- US Nationwide Distribution in the states of AZ,CA, NC, NJ, NY, PA, SC, TX, and WA. There were no OUS consignees.
- Lot, serial or model codes
- Lot/Serial numbers: B180228-751 through B180228-779;Model/Product Code: 359100;the expiration date of these devices is 2021-06-05.
- 510(k) numbers
- ["K130498"]
- PMA numbers
- not on file
- Product code
- MBP
- Firm FEI
- 3005031160
- Firm city
- Belgrade
- Firm state
- MT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28