Cyberonics, Inc: VNS Therapy DemiPulse Generator Model 103 and VNS Therapy DemiPulse Duo Generator Model 104; CYBERONICS, INC.

Recall number
Z-1526-2015
Recalling firm
Cyberonics, Inc
Product
VNS Therapy DemiPulse Generator Model 103 and VNS Therapy DemiPulse Duo Generator Model 104; CYBERONICS, INC.
Status
Terminated
Initiated
2015-03-06
Posted
2015-04-27
Terminated
2015-10-09
Reason
The pulse generators have a lower battery longevity than specified in their design requirement as a result of the devices being inadvertently left in a programmed ON state during manufacture.
Root cause
Nonconforming Material/Component
Action
On 03/6/2015, the recalling firm sent a recall notification letter to their consignees. The letter included an effectiveness card which is to be signed by the consignee and returned to the recalling firm.
Quantity
13 units
Distribution
AR, TX, GA, CA, MI, IN, FL, AZ, NV
Lot, serial or model codes
Serial Numbers: 37627, 38473, 38539, 41446, 42887, 43491, 44186, 44339, 44391, 44443, 44455, 44510, 50715
510(k) numbers
not on file
PMA numbers
not on file
Product code
LYJ
Firm FEI
1000117004
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register