FDA Medical Device Recalls (openFDA)
Coloplast Manufacturing US, LLC: Self-Cath Pediatric Catheter Size CH 06, Product # 5044101400, Catakig # 306. Product Usage: The Self-Cath intermittent urinary catheter discussed in this document is a pediatric size CH06 French catheter. The catheter is indicated for use by patients with chronic urine retention and patients with a post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing the urine to drain.
- Recall number
- Z-1525-2016
- Recalling firm
- Coloplast Manufacturing US, LLC
- Product
- Self-Cath Pediatric Catheter Size CH 06, Product # 5044101400, Catakig # 306. Product Usage: The Self-Cath intermittent urinary catheter discussed in this document is a pediatric size CH06 French catheter. The catheter is indicated for use by patients with chronic urine retention and patients with a post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing the urine to drain.
- Status
- Terminated
- Initiated
- 2016-03-22
- Posted
- not on file
- Terminated
- 2016-10-11
- Reason
- Coloplast Corp. is voluntarily recalling a single lot of Self-Cath Pediatric French Size CH 06 catheters because a small portion of the lot was identified to have missing eyelets. This recall is based upon a complaint Coloplast received regarding a catheter malfunction.
- Root cause
- Nonconforming Material/Component
- Action
- Coloplast sent an "Urgent Medical Device Recall" letter dated March 22, 2016 to customers. The letter described the problem and the product involved in the recall. Advised consignees to immediately examine, quarantine and return the affected product. If the product has been further distributed, consignees should identify their customers and notify them of the recall. It also requested consignees to complete and return the Acknowledgement Form. For questions contact Coloplast Customer Service directly at 800-533-0464.
- Quantity
- 31,320
- Distribution
- US Nationwide Distribution in the states of: AL, AR, C, FL, IL KY, LA, MD, MA, MI, MN, MO, NJ,NY, NC, OK,OR, PA, TN, TX, VA and WA.
- Lot, serial or model codes
- 4464131
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- EZL
- Firm FEI
- 2125050
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as