FDA Medical Device Recalls (openFDA)
Sofradim Production: Versatex Monofilament Mesh 50 x 50cm, Product No. VTX5050M The device is made out of a macroporous three-dimensional monofilament polyester textile. Its intended use is for the repair of abdominal wall hernias or other fascial deficiencies that require the addition of a reinforcing material.
- Recall number
- Z-1523-2018
- Recalling firm
- Sofradim Production
- Product
- Versatex Monofilament Mesh 50 x 50cm, Product No. VTX5050M The device is made out of a macroporous three-dimensional monofilament polyester textile. Its intended use is for the repair of abdominal wall hernias or other fascial deficiencies that require the addition of a reinforcing material.
- Status
- Terminated
- Initiated
- 2018-02-23
- Posted
- 2018-04-26
- Terminated
- 2020-05-12
- Reason
- There have been patient reports of abdominal hernia recurrence following hernia repair. The recalling firm has added a statement to the device's Instructions for Use to make surgeons aware of risks in using the device in Transversus Abdominis Muscle Release (TAR) procedures.
- Root cause
- Other
- Action
- All consignees were notified via Federal Express or certified mail on February 23, 2018. Customers were advised of the added warning on the product labeling. Customers were requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice. Customers with questions were directed to contact their Medtronic Representative, or Customer Service at 800-962-9888, option 2.
- Quantity
- 703
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution to the states of: CA, CT, FL, GA, IL, IN, KY, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, PA, TN, TX, VA, WA and WI, and to the countries of : Belgium, Finland, France, Germany, Ireland, Israel, Italy, Netherlands, Russian Federation, Spain, Switzerland, and UK.
- Lot, serial or model codes
- UDI - 10884521532298 Lots SPC1091X SPJ0706X SPJ1016X SQA0990X SQI0302X SQJ0443X SQK0319X SRA2188X SRI0548X
- 510(k) numbers
- ["K150091"]
- PMA numbers
- not on file
- Product code
- FTL
- Firm FEI
- 3003521774
- Firm city
- Trevoux
- Firm state
- not on file
- Firm country
- France
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28