Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): CareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (CIED).

Recall number
Z-1522-2021
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Product
CareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (CIED).
Status
Open, Classified
Initiated
2021-04-06
Posted
not on file
Terminated
not on file
Reason
There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).
Root cause
Software design
Action
Beginning 06-APR-2021, Medtronic notified risk managers and practice managers of the field corrective action notification via 2-day UPS delivery of the consignee letter. Confirmation delivery of the consignee letter will be obtained. Medtronic Field Representatives have been instructed to begin installation of the software update. Outside the United States (OUS), communication began on 06- APR-2021 using locally approved methods. Medtronic Field Representatives will begin installation of the software update as it is available locally. Consignees were advised that Medtronic representatives can assist in installing the software update on programmers. During device interrogation with an updated programmer at implant or the next regularly scheduled clinic visit, a patients device will automatically receive the updated software. A courtesy notification will be provided to physicians. No confirmation will be required for the courtesy notification.
Quantity
16,311 devices
Distribution
Worldwide
Lot, serial or model codes
All serial numbers
510(k) numbers
not on file
PMA numbers
["P890003S386"]
Product code
OSR
Firm FEI
2182208
Firm city
Mounds View
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register