FDA Medical Device Recalls (openFDA)
Abbott Laboratories: CELL-DYN Emerald Cleaner, product number 09H46-02. 960mL. IVD; Hematology: The CELL-DYN Emerald Cleaner is designed for use on the CELL-DYN Emerald System, which is an automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories. The CELL-DYN Emerald Cleaner is an enzymatic cleaner used to clean the measurement system and the fluidics.
- Recall number
- Z-1522-2016
- Recalling firm
- Abbott Laboratories
- Product
- CELL-DYN Emerald Cleaner, product number 09H46-02. 960mL. IVD; Hematology: The CELL-DYN Emerald Cleaner is designed for use on the CELL-DYN Emerald System, which is an automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories. The CELL-DYN Emerald Cleaner is an enzymatic cleaner used to clean the measurement system and the fluidics.
- Status
- Terminated
- Initiated
- 2016-03-24
- Posted
- 2016-04-19
- Terminated
- 2017-01-04
- Reason
- Three lots of CELL-DYN Emerald Cleaner may have occurrences where the analyzer generates Quality Control Out of Range Low for parameters RBC and PLT.
- Root cause
- Under Investigation by firm
- Action
- All customers who were shipped the affected lots were sent a Product Recall letter on March 24, 2016 via Federal Express. Letters were titled Product Recall - Immediate Action Required. An additional four lots were discovered to be affected by the recall and the firm send letters to customers affected by these lots on June 27, 2016
- Quantity
- 10,364 plus 9410 additional units for recall expansion.
- Distribution
- Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Austria, Bosnia & Herze, Cameroon, Egypt, Germany, Indonesia, Iraq, Lebanon, Mali, Montenegro, Netherlands, Philippines, Tanzania, Thailand, UAE, New Zealand, Malaysia, Mexico, South Korea, Jamaica, Turks & Caicos, Italy, Canada, Vietnam, Singapore, Argentina, Colombia, France, Paraguay, Australia, Uruguay, Bahrain, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Ethiopia, Finland, Greece, Switzerland, United Kingdom, Brazil, Hong, Kong, Hungary, India, Ireland, Israel, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Turkey, Jamaica, St. Martin, Nigeria, Oman, Pakistan, Philippines, Poland, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovakia, Slovenia, Spain, Guatemala, Peru, and Uruguay.
- Lot, serial or model codes
- Lot 6853, Expiration date: 31 JUL2017; Lot 6901, Expiration date: 31JUL2017; Lot 6953, Expiration date: 30SEP2017; May 4, 2016 - Firm advised SAN-DO that it has expanded the recall to include three additional lots: Lot 6991, Expiration date: 30NOV2017; Lot 7024, Expiration date: 31DEC2017; Lot: 7027, Expiration date: 31DEC2017. June 28, 2016 - firm added an additional 4 lots to recall: Lot 7044, expiration date 31JAN2018; Lot 7082, expiration date 28FEB2018; Lot 7110, expiration date 31MAR2018; Lot 7119, expiration date 30APR2018.
- 510(k) numbers
- ["K081495"]
- PMA numbers
- not on file
- Product code
- GKZ
- Firm FEI
- 2919069
- Firm city
- Santa Clara
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28