Skip to the content

FDA Medical Device Recalls (openFDA)

SonarMed Inc: AW-S035 NEO SONARMED SENSOR 3.5MM. Used in conjunction with SonarMed Monitor for airway monitoring.

Recall number
Z-1521-2024
Recalling firm
SonarMed Inc
Product
AW-S035 NEO SONARMED SENSOR 3.5MM. Used in conjunction with SonarMed Monitor for airway monitoring.
Status
Open, Classified
Initiated
2024-03-21
Posted
2024-04-18
Terminated
not on file
Reason
The failure to detect the partial obstruction in a 2.5mm sensor.
Root cause
Process control
Action
An Urgent Medical Device Recall notification dated 3/25/2024 was mailed to consignees via UPS notifying them of this SonarMed Sensor recall. Consignees were asked to discontinue use of affected devices, quarantine any affected devices, and return all quarantined devices to Medtronic by obtaining a Returned Goods Authorization (RGA) from rs.covidienfeedbackcustomerservice@medtronic.com. If product was purchased through a distributor the distributor can arrange for return. The provided notification is to be provided to those within consignee organizations that need to be made aware and if product was further distributed. Consignees with questions are to contact customer service at 800-962-9888, Option 2.
Quantity
500 units
Distribution
Domestic: AK, AZ, CA, FL, IL, KY, MD, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, & WA.
Lot, serial or model codes
Product Number/CFN: AW-S035; UDI-DI: 10851334007206; Serial Numbers: A20221012, A20221019, A20221026, A20221109, A20221110, A20221111, A20221114, A20221212, A20221213, A20221214, A20221215, A20221216, A20230110, A20230111, A20230112, A20230113, A20230116, A20230130, A20230131, A20230202, A20230206, A20230306, A20230307.
510(k) numbers
["K193058"]
PMA numbers
not on file
Product code
OQU
Firm FEI
3009923584
Firm city
Carmel
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register