Ortho-Clinical Diagnostics: Software Version 2.8 & Below on VITROS 5,1 FS Chemistry Systems, Catalog Number 6801375, Global Trade Item Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Global Trade Item Number 10758750001644; IVD. Intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides and VITROS Chemistry Products MicroTip Reagents.
- Recall number
- Z-1521-2015
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- Software Version 2.8 & Below on VITROS 5,1 FS Chemistry Systems, Catalog Number 6801375, Global Trade Item Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Global Trade Item Number 10758750001644; IVD. Intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides and VITROS Chemistry Products MicroTip Reagents.
- Status
- Terminated
- Initiated
- 2015-03-10
- Posted
- 2015-04-25
- Terminated
- 2018-06-01
- Reason
- Software Anomaly; It is possible for the device to process samples with a cartridge other than the intended cartridge, potentially leading to erroneous patient results. If this anomaly occurred previously, an indication would be a series of results that were similar across multiple samples (i.e., results could be believable or outside of the Reportable Range for the intended assay).
- Root cause
- Software in the Use Environment
- Action
- On 3/10/2015, URGENT PRODUCT CORRECTION NOTIFICATION letters (Ref. CL2015-047, dated 3/10/2015) were sent to users notifying them of the issue and advising them of the required actions. Customers are to complete and return the Confirmation of Receipt no later than March 23, 2015. The firm is currently working on a resolution and will issue a follow up notification upon the software's availability. For questions, please contact Customer Technical Services at 1-800-421-3311.
- Quantity
- VITROS 5.1 System: Domestic - 909; Foreign - 1250; 5,1 Refurbished: Domestic - 66 units, Foreign - 152
- Distribution
- Worldwide Distribution -- USA, Puerto Rico, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.
- Lot, serial or model codes
- Software Version 2.8 & Below; VITROS 5,1 FS Chemistry System Serial Numbers 34000119 - 34002323; VITROS 5,1 Refurbished: Serial Numbers - Domestic S/Ns: 34000133, 34000193, 34000197, 34000212, 34000285, 34000316, 34000318, 34000337, 34000360, 34000420, 34000451, 34000464, 34000478, 34000495, 34000497, 34000529, 34000538, 34000552, 34000557, 34000590, 34000596, 34000601, 34000631, 34000642, 34000761, 34000799, 34000805, 34000833, 34000851, 34000893, 34000946, 34000947, 34001006, 34001172, 34001218, 34001240, 34001243, 34001248, 34001251, 34001274, 34001297, 34001311, 34001317, 34001345, 34001378, 34001379, 34001389, 34001390, 34001396, 34001422, 34001472, 34001504, 34001569, 34001572, 34001589, 34001610, 34001612, 34001626, 34001633, 34001668, 34001755, 34001809, 34001840, 34001850, 34001875, 34001897; Foreign S/Ns: 34000123, 34000143, 34000191, 34000215, 34000219, 34000235, 34000251, 34000256, 34000260, 34000261, 34000319, 34000331, 34000380, 34000382, 34000406, 34000408, 34000412, 34000426, 34000431, 34000454, 34000464, 34000473, 34000501, 34000523, 34000530, 34000541, 34000542, 34000546, 34000548, 34000553, 34000558, 34000576, 34000577, 34000584, 34000585, 34000609, 34000612, 34000615, 34000616, 34000617, 34000618, 34000620, 34000625, 34000637, 34000644, 34000672, 34000744, 34000745, 34000755, 34000759, 34000776, 34000793, 34000800, 34000814, 34000831, 34000844, 34000846, 34000852, 34000856, 34000870, 34000876, 34000887, 34000892, 34000902, 34000912, 34000920, 34000938, 34000956, 34000969, 34000976, 34001007, 34001009, 34001044, 34001045, 34001071, 34001075, 34001098, 34001099, 34001111, 34001129, 34001154, 34001161, 34001171, 34001182, 34001204, 34001208, 34001212, 34001234, 34001247, 34001254, 34001271, 34001278, 34001284, 34001288, 34001289, 34001293, 34001298, 34001301, 34001316, 34001325, 34001337, 34001340, 34001348, 34001367, 34001369, 34001399, 34001400, 34001407, 34001410, 34001416, 34001423, 34001427, 34001429, 34001457, 34001490, 34001518, 34001526, 34001551, 34001573, 34001580, 34001581, 34001582, 34001587, 34001589, 34001594, 34001610, 34001618, 34001623, 34001627, 34001673, 34001688, 34001693, 34001725, 34001738, 34001790, 34001796, 34001810, 34001822, 34001831, 34001833, 34001849, 34001855, 34001859, 34001860, 34001867, 34001871, 34001872, 34001881, 34001905, 34001916, 34001923, 34001995. --- PLEASE NOTE: The serial numbers are sequential, but not all analyzers are released for distribution. Therefore, the amount of each system manufactured may not equal the amount of each system distributed.
- 510(k) numbers
- ["K031924","K063144"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28