Applied Medical Resources Corp: Applied Medicals Epix and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and manipulating tissue during general or laparoscopic surgery.

Recall number
Z-1521-2013
Recalling firm
Applied Medical Resources Corp
Product
Applied Medicals Epix and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and manipulating tissue during general or laparoscopic surgery.
Status
Terminated
Initiated
2013-05-01
Posted
2013-06-07
Terminated
2014-02-06
Reason
Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a potential for the trigger to fracture and the jaws to remain in the closed position. The likelihood of this situation to occur and result in permanent patient injury is highly unlikely;
Root cause
Nonconforming Material/Component
Action
Applied Medical notified customers of the recall beginning on May 13, 2013. The "URGENT: MEDICAL DEVICE RECALL" letter identified the product, problem, and actions to be taken by the customers. For questions regarding this notice, contact the firm at (949) 713-8041.
Quantity
127,305 units
Distribution
Worldwide Distribution: USA Nationwide and countries of: Argentina, Austria, Australia, Bahrain, Belgium, Canada, Chile, Columbia, Denmark, Ecuador, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Jordan, Kuwait, Malaysia, Lebanon, Lybia, Mexico, Netherlands, New Zealand, Pakistan, Peru, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Spain, South Africa, Switzerland, Taiwan-Province of China,Turkey, United Arabs Emirates United Kingdom, Uruguay, and Venezula-Bolivarian Republic.
Lot, serial or model codes
Model Numbers C4130 and C4140.  MODEL NUMBERS    C4130- 5mm x 35cm Epix/Direct Drive Laparoscopic Grasper  LOT NUMBERS  1145332, 1145943, 1145944, 1145945, 1145992, 1145993, 1145994, 1146371, 1147900, 1147901, 1147955, 1148172, 1148173, 1148174, 1148180, 1148181, 1148184, 1148190, 1148191, 1148192, 1148194, 1148201, 1148203, 1148677, 1148765, 1148766, 1148767, 1148835, 1149040, 1149041, 1149690, 1149757, 1150168, 1150169, 1150170, 1150490, 1150634, 1151246, 1151247, 1151248, 1152274, 1152728, 1153332, 1154394, 1154637,  1155135, 1155473, 1155948, 1156004, 1156010, 1156011, 1156012, 1156270, 1156360, 1156452, 1156577, 1156857, 1157218, 1157679, 1157950, 1157951, 1157952, 1157953, 1157954, 1158052, 1158131, 1158216, 1158217, 1158425, 1159491, 1160167, 1161136, 1161137, 1161398, 1161638, 1162112, 1162503, 1162504, 1162588, 1163399, 1163558, 1163559, 1163575, 1164580, 1164581, 1164582, 1164583, 1168124, 1168637, 1168638, 1169117, 1171168, 1171631, 1172170, 1172668, 1173153, 1174995, 1174996, 1176616, 1177407, 1177408, 1178109, 1180295, 1181554, 1182940, 1183774, 1184416, 1184959, 1187241, 1188110, 1188745, 1189773, 1191368, 1191809.  C4140- 5mm x 45cm Epix/Direct Drive Laparoscopic  Grasper  1145559, 1147851, 1147852, 1147960, 1148210, 1148211, 1149038, 1149039, 1149689, 1149936, 1150488, 1150489, 1151249, 1152594, 1153330, 1155645, 1157217, 1157949, 1158644, 1160033, 1160170, 1162111, 1167871, 1169855, 1171497, 1172009, 1173587, 1174121, 1177578, 1179347, 1180753, 1181346, 1182924, 1183769, 1184961, 1188111, 1189182  Various Kits Containing Either C4130 or C4140 1148079, 1148139, 1148145, 1148250, 1148514, 1148770, 1148780, 1149100, 1149605, 1150455, 1151831, 1153074, 1153702, 1154931, 1154997, 1155332, 1156581, 1156812, 1157090, 1158535, 1160152, 1160227, 1160388, 1161766, 1162189, 1162343, 1163348, 1163881, 1164590, 1164796, 1164831, 1165443, 1166520, 1168227, 1169190, 1169456, 1169620, 1169772, 1170151, 1170378, 1171280, 1171878, 1171879, 1172330, 1173319, 1173903, 1174093, 1175987, 1175991, 1176171, 1178210, 1178498, 1178581, 1179376, 1182679, 1182972, 1183303, 1184673, 1185599, 1185762, 1186257, 1186868, 1187082, 1188827, 1189581, 1190574, 1191055, 1193112, 1193147, 1193648, 1193688, 1194964,  1194965
510(k) numbers
not on file
PMA numbers
not on file
Product code
NWV
Firm FEI
2027111
Firm city
Rancho Santa Margarita
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register