Skip to the content

FDA Medical Device Recalls (openFDA)

Datex - Ohmeda, Inc: GE Healthcare Innova 3100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System)

Recall number
Z-1521-2008
Recalling firm
Datex - Ohmeda, Inc
Product
GE Healthcare Innova 3100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System)
Status
Terminated
Initiated
2008-05-19
Posted
2008-09-30
Terminated
2016-06-30
Reason
GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room).
Root cause
Software design
Action
An Urgent Medical Device Correction letter, dated June 2008 was sent to customers The letter described the issue, affected products and provided short term user recommendations. GE developed a new software release that corrects this condition and will launch the corrective action to upgrade all affected systems.
Quantity
107
Distribution
Worldwide Distribution - AL, AZ, CA, FL,, GA, HI, IA,IL, IN, KY, LA,MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV, Puerto Rico. OUS to include: Australia, Byelarus, China, Denmark, France, Germany, Greece, Honduras, Iran, Iraq, Italy, Japan, Latvia, Lebanon, Mexico, Poland, Russia, Saudi Arabia, Turkey, United Kingdom, and Venezuela.
Lot, serial or model codes
Systems Id: 8613100, 0910163013, 4103323100, 4103623100, 5034353100, 7083333100, 7278253100, 8019643100, 82416040010, 082416040016, 082416040017, 082416050002, 082416070004, 082416100008, 82416120005, 082416120006, 082416130013, 0002581738X, 06074VAS01, 201541LAB2, 214324INNOVA2, 24516572980IN, 309691P3100, 310423INNOVA1, 313593IC1, 313593IC2, 314525SACL2, 314768SMCL3, 318448INNOVA2, 318449INNOVA2, 409899LAB3, 413447CATH, 413540IN1, 417269SCLD, 502587LAB4, 518525LAB1, 540953INNOVA, 561266IN31, 563421CATHLAB8, 573815BCR1, 600130VA01, 601553CL3, 602239LAB4, 602239LAB6, 602266LAB2, 603778ECL1, 608524NOVA, 615338CV2, 615769SKYCV2, 619340VA8, 619482INNOVA1, 619541INNOVA31, 619552INNOVA1, 702616INNOVA, 7028803100A, 713794LAB7, 713794LAB9, 714966INNOVA, 718283INNOVA3, 724773C1, 731422CV1, 732294INOV, 765453INNOVA1, 770991CL3, 773792CL3100, 773NORWG3100, 803329CL31, 808547QCV2A, 812238LIB1, 812283IN31, 812482INNOVA, 815599FHN3100, 817433HSW3100, 83INNOVA1188379, 850678IN3100, 904202CATH3, 908522LAB1, 928779INNOVA1, 936266CL1, 956389INN31, 956698INN31, 956994INN1, 956994INN2, 973971LAB3, A5108239, A5203215, BPINNOVA2, BPINNOVA3R, DK1066VA01, FHORLCL1, H1420VAS04, HC1536XR04, IQ1083VA01, IR1747VA01, LB1160XR07, LV4012VA01, M2848425, M2970614, PL1622VA01, RU1156VA01, RU1482VA02, RU1550VA01, RU2457VA01, TT-2586581-CO, YV0022, YV0026, YV0036.
510(k) numbers
["K031637","K042053"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2183066
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register