Diamedix Corporation: Product is labeled in part: "***is immunosimplicity***HSV 1&2 IgM***Enzyme Immunoassay Test Kit***For the Detection of HSV 1&2 IgM antibodies***Catalog No: 720-350***Exp. Date:***Contains Sufficient for 96 Tests***C - Corrosive Harmful***Lot No: ***For In Vitro Diagnostic Use***Caution, consult accompanying documents.***Manufacturer***Diamedix***A Subsidiary of IVAX Diagnostics, Inc. MIAMI, FL 33127 USA-1-800-327-4565***Biohazard***" The Is-HSV 1&2 IgM test kits are packaged in cardboard boxes with individual components secured within a foam insert. The Diamedix Is-HSV 1&2 IgM Test Kit is intended for the qualitative detection of IgM antibodies to herpes simplex virus (HSV) type 1 and/or type 2 in human serum by indirect enzyme immunoassay. This test can aid in the diagnosis of a primary or reactived infection with HSV. The performance of this assay has not bee established for use in neonates, infants, or on cord blood, and immunocompromised patients.
- Recall number
- Z-1520-2012
- Recalling firm
- Diamedix Corporation
- Product
- Product is labeled in part: "***is immunosimplicity***HSV 1&2 IgM***Enzyme Immunoassay Test Kit***For the Detection of HSV 1&2 IgM antibodies***Catalog No: 720-350***Exp. Date:***Contains Sufficient for 96 Tests***C - Corrosive Harmful***Lot No: ***For In Vitro Diagnostic Use***Caution, consult accompanying documents.***Manufacturer***Diamedix***A Subsidiary of IVAX Diagnostics, Inc. MIAMI, FL 33127 USA-1-800-327-4565***Biohazard***" The Is-HSV 1&2 IgM test kits are packaged in cardboard boxes with individual components secured within a foam insert. The Diamedix Is-HSV 1&2 IgM Test Kit is intended for the qualitative detection of IgM antibodies to herpes simplex virus (HSV) type 1 and/or type 2 in human serum by indirect enzyme immunoassay. This test can aid in the diagnosis of a primary or reactived infection with HSV. The performance of this assay has not bee established for use in neonates, infants, or on cord blood, and immunocompromised patients.
- Status
- Terminated
- Initiated
- 2012-01-16
- Posted
- 2012-05-10
- Terminated
- 2012-05-18
- Reason
- On 01/16/20112 Diamedix initiated a recall of IS-HSV 1&2 IgM Test Kit, Catalog # 720-350, Lot 60701, Exp. June 30, 2012. The Positive Control for this lot is Out of Specification.
- Root cause
- Nonconforming Material/Component
- Action
- Diamedix sent Recall notification letters dated January 16, 2012, to all affected customers by Certified Mail. Consignees were asked to review their inventory and then contact Diamedix to request replacement product. Customers were instructed to call 305-324-2314 with questions. For questions regarding this recall call 305-324-2560.
- Quantity
- 442 kits
- Distribution
- Nationwide Distribution and Puerto Rico.
- Lot, serial or model codes
- Lot # 60701
- 510(k) numbers
- ["K002262"]
- PMA numbers
- not on file
- Product code
- LGC
- Firm FEI
- 1044713
- Firm city
- Miami
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28