Zimmer Inc.: Zimmer NexGen Complete Knee Solution Legacy Knee - Posterior Stabilized LPS-Flex Femoral Component Precoat, Size D left , sterile, Zimmer, Warsaw, IN; REF 00-5960-014-51.

Recall number
Z-1520-2009
Recalling firm
Zimmer Inc.
Product
Zimmer NexGen Complete Knee Solution Legacy Knee - Posterior Stabilized LPS-Flex Femoral Component Precoat, Size D left , sterile, Zimmer, Warsaw, IN; REF 00-5960-014-51.
Status
Terminated
Initiated
2009-04-17
Posted
2009-06-25
Terminated
2009-11-12
Reason
The implant surface may not have been polished adequately, which could result in increased wear and polyethylene debris generation.
Root cause
Employee error
Action
Consignees were notified by phone beginning on 4/17/09 and email and letter notifications were sent on 4/22/09. A dear surgeon letter entitled Urgent: Medical Device Recall, dated April 28, 2009, is being sent to implanting surgeons recommending patient monitoring. Consignees are to return the response form indicating their receipt of the notification. Questions or concerns should be directed to (800) 846-4637 or www.zimmer.com.
Quantity
5
Distribution
Worldwide Distribution -- USA including states of Arizona, California, Colorado, Hawaii, Illinois, Indiana, Michigan, Minnesota, Mississippi, New Jersey, New Mexico, New York, Ohio, Oregon, and Pennsylvania and country of Canada.
Lot, serial or model codes
Lot 61208295.
510(k) numbers
not on file
PMA numbers
["P060037"]
Product code
NJL
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register