Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Corporation: Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test

Recall number
Z-1519-2018
Recalling firm
Roche Diagnostics Corporation
Product
Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test
Status
Terminated
Initiated
2018-01-02
Posted
not on file
Terminated
2019-04-25
Reason
The previous version of method sheets for the Acetaminophen assay on the COBAS INTEGRA 400 plus/800/800 CTS analyzers and the MODULAR ANALYTICS P module are missing the following statement in the Limitations-interference section: "In very rare cases, gammopathy, in particular type IgM (Waldenstrm s macroglobulinemia), may cause unreliable results."
Root cause
Error in labeling
Action
On January 2, 2018 an URGENT MEDICAL DEVICE CORRECTION letter was issued to customers with notification to the issue and information regarding the updated method sheets and that they were available for use. The COBAS INTEGRA 400 plus/800/800 CTS was updated from affected method sheet version 3.0 to 4.0. The MODULAR ANALYTICS P was updated from affected method sheet version 4.0 to 5.0 No product return is required. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Quantity
35,111
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot codes: 11021500 7/31/2016 12361201 5/31/2017 11031700 9/30/2018 15247301 8/31/2017 11071400 1/31/2016 18036001 11/30/2017 11071600 8/10/2016 19329601 1/31/2018 11101500 1/21/2016 20511301 5/31/2018 11111600 4/30/2018 24076801 8/31/2018 60482201 2/29/2016 60852601 5/31/2016 61385701 9/30/2016 62032101 1/31/2017
510(k) numbers
["K110726"]
PMA numbers
not on file
Product code
LDP
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register