Skip to the content

FDA Medical Device Recalls (openFDA)

Omnia SRL: Omnia Soft Tissue Punch is a dental instrument designed to remove with precision a circular section of oral soft tissue (e.g., gum, palatal mucosa) during dental surgery, typically to allow access to a site for drilling and dental/orthodontic implant placement without the use of a scalpel. Model: 32.Z2000.00, 32.Z2002.00, 32.Z2004.00, 32.Z2006.00, 32.Z2007.00, 32.Z2008.00, 32.Z2009.00, 32.Z2020.00, 32.Z2022.00, 32.Z2024.00, 32.Z2026.00, 32.Z2027.00, 32.Z2028.00, 32.Z2030.00, 32.Z2032.00, 32.Z2034.00, 32.Z2036.00, 32.Z2038.00, 32.Z2040.00, 72.T1950.00, 72.T1951.00, 72.T1952.00, 72.T1953.00

Recall number
Z-1518-2021
Recalling firm
Omnia SRL
Product
Omnia Soft Tissue Punch is a dental instrument designed to remove with precision a circular section of oral soft tissue (e.g., gum, palatal mucosa) during dental surgery, typically to allow access to a site for drilling and dental/orthodontic implant placement without the use of a scalpel. Model: 32.Z2000.00, 32.Z2002.00, 32.Z2004.00, 32.Z2006.00, 32.Z2007.00, 32.Z2008.00, 32.Z2009.00, 32.Z2020.00, 32.Z2022.00, 32.Z2024.00, 32.Z2026.00, 32.Z2027.00, 32.Z2028.00, 32.Z2030.00, 32.Z2032.00, 32.Z2034.00, 32.Z2036.00, 32.Z2038.00, 32.Z2040.00, 72.T1950.00, 72.T1951.00, 72.T1952.00, 72.T1953.00
Status
Terminated
Initiated
2021-03-30
Posted
2021-04-29
Terminated
2023-11-08
Reason
Omnia SRL received a notification from the contract manufacturer informing them of inadequate sterilization processes which have been taking place at their contracted sterilizing facility since 2016.
Root cause
Vendor change control
Action
On 03/30/2021, Omnia issued an Urgent Medical Device Removal notice via letter to customers notifying them Omnia has been made aware of a situation at a contract sterilization facility regarding inadequate or inappropriate sterilization processes being followed by employees of the contract sterilizer.
Quantity
276,901 units
Distribution
Domestic: AK, IL, PA, OK, TX, UT, NY, CA. Foreign: Australia, Austria, Belgium, Bulgaria, Denmark, Dominican Republic, Estonia, France, Germany, Greece, Hungary, Ireland, Italy, Lebanon, Netherlands, New Zealand, Poland, Russia, Slovenia, Spain, Sweden, Switzerland, United Kingdom, Uzbekistan, Canada, Malta,
Lot, serial or model codes
Lot:118, 317, 319, 516, 616, 617, 618, 620, 716, 719, 720, 817, 818, 916, 920, 1016, 1018, 1216, 1217, 1316, 1317, 1318, 1319, 1418, 1516, 1618, 1718, 1720, 1816, 1817, 1918, 1919, 2018, 2117, 2118, 2119, 2216, 2316, 2319, 2416, 2417, 2619, 2713, 2717, 2718, 2719, 2816,2818, 2819, 2916, 3218, 3219, 3416, 3517,3716, 3719, 3816, 3818, 3916, 3917, 3918, 3919, 4118, 4217, 4220, 4318, 4320, 4416, 4417, 4419, 4617, 4715, 4716, 4816, 4819, 4820, 4915, 4920, 5118, 5319, 0317, 190B, 1920A, 1920B, 2518A, 2518B, 4820A
510(k) numbers
not on file
PMA numbers
not on file
Product code
EME
Firm FEI
3004361903
Firm city
Fidenza
Firm state
not on file
Firm country
Italy

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register