Skip to the content

FDA Medical Device Recalls (openFDA)

Datex - Ohmeda, Inc: GE Healthcare Innova 2100 IQ Digital Fluoroscopic Imaging System (Versatile cardiac and Vascular imaging system). The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy of cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. These devices are not intended for mammography applications.

Recall number
Z-1518-2008
Recalling firm
Datex - Ohmeda, Inc
Product
GE Healthcare Innova 2100 IQ Digital Fluoroscopic Imaging System (Versatile cardiac and Vascular imaging system). The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy of cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. These devices are not intended for mammography applications.
Status
Terminated
Initiated
2006-08-17
Posted
2008-09-30
Terminated
2009-06-09
Reason
User was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to incorrect overall dose management, resulting in unnecessary exposure and can be a contributing factor for skin injury. The alarm can be heard in the control room.
Root cause
Radiation Control for Health and Safety Act
Action
GE Field Engineers are visiting all affected customers to apply updated software. For assistance contact Datax-Ohmeda, Inc., at 1-608-221-1551.
Quantity
281
Distribution
within US: AL, AR, AZ, CA,CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico. OUS to include: Algeria, Australia, Belgium, Brazil, Canada, China, Columbia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Korea, Kuwait, Latvia, Malaysia, Mexico, New Zealand, Norway, Russia, Saudi Arabia, Singapore, Spain, Switzerland, Syria, Taiwan, Turkey, and United Kingdom
Lot, serial or model codes
serial numbers: 00000007C20041, 00000046457BU0, 00000046582BU1, 00000051324BU5, 000000MI000555, 00000454582BU8, 00000455797BU9, 00000459650BU8, 00000459651BU6, 00000459652BU4, 00000459653BU2, 00000459654BU0, 00000459655BU7, 00000459656BU5, 00000459657BU3, 00000459658BU1, 00000459659BU9, 00000459661BU5, 00000459663BU1, 00000459664BU9, 00000459665BU6, 00000459667BU2, 00000459668BU0, 00000459669BU8, 00000459670BU6, 00000459672BU2, 00000459673BU0, 00000459674BU8, 00000459675BU5, 00000459676BU3, 00000459677BU1, 00000459891BU8, 00000464564BU4, 00000464565BU1, 00000464566BU9, 00000464567BU7, 00000464568BU5, 00000464570BU1, 00000464571BU9, 00000464572BU7, 00000464573BU5, 00000464574BU3, 00000464576BU8, 00000464577BU6, 00000464578BU4, 00000464579BU2, 00000464580BU0, 00000464581BU8, 00000464582BU6, 00000464585BU9, 00000464586BU7, 00000464587BU5, 00000464588BU3, 00000464589BU1, 00000464590BU9, 00000464591BU7, 00000464592BU5, 00000464593BU3, 00000465560BU1, 00000465794BU6, 00000465795BU3, 00000465796BU1, 00000465821BU7, 00000465822BU5, 00000465823BU3, 00000465825BU8, 00000465827BU4, 00000465828BU2, 00000465829BU0, 00000465830BU8, 00000465831BU6, 00000465832BU4, 00000465834BU0, 00000465835BU7, 00000465836BU5, 00000465837BU3, 00000465838BU1, 00000465839BU9, 00000465840BU7, 00000465841BU5, 00000465842BU3, 00000465843BU1, 00000465845BU6, 00000465846BU4, 00000465848BU0, 00000465849BU8, 00000466836BU4, 00000466838BU0, 00000466839BU8, 00000466841BU4, 00000466843BU0, 00000466844BU8, 00000466845BU5, 00000466846BU3, 00000466847BU1, 00000466848BU9, 00000466849BU7, 00000466850BU5, 00000466851BU3, 00000466852BU1, 00000466853BU9, 00000466856BU2, 00000466857BU0, 00000466859BU6, 00000466860BU4, 00000466861BU2, 00000466862BU0, 00000466863BU8, 00000466864BU6, 00000466865BU3, 00000478319BU7, 00000478320BU5, 00000478323BU9, 00000478324BU7, 00000478326BU2, 00000478327BU0, 00000478328BU8, 00000478330BU4, 00000478331BU2, 00000478333BU8, 00000478337BU9, 00000478338BU7, 00000478339BU5, 00000478340BU3, 00000478341BU1, 00000478342BU9, 00000478345BU2, 00000478346BU0, 00000478347BU8, 00000478348BU6, 00000485080BU6, 00000489536BU3, 00000489537BU1, 00000489538BU9, 00000489539BU7, 00000489540BU5, 00000489542BU1, 00000489543BU9, 00000489544BU7, 00000489546BU2, 00000489547BU0, 00000489548BU8, 00000489549BU6, 00000489551BU2, 00000489553BU8, 00000489555BU3, 00000489557BU9, 00000489558BU7, 00000489561BU1, 00000489562BU9, 00000489564BU5, 00000489565BU2, 00000489683BU3, 00000489684BU1, 00000489688BU2, 00000489738BU5, 00000489744BU3, 00000489747BU6, 00000489753BU4, 00000489756BU7, 00000489757BU5, 00000489761BU7, 00000489763BU3, 00000489770BU8, 00000489772BU4, 00000498661BU8, 00000498679BU0, 00000498710BU3, 00000498711BU1, 00000498712BU9, 00000498713BU7, 00000498715BU2, 00000498717BU8, 00000498718BU6, 00000498719BU4, 00000498720BU2, 00000498721BU0, 00000498722BU8, 00000498723BU6, 00000508755BU6, 00000509439BU6, 00000512697BU4, 00000512701BU4, 00000512713BU9, 00000512715BU4, 00000512727BU9, 00000512818BU6, 00000512819BU4, 00000512821BU0, 00000512822BU8, 00000512823BU6, 00000512828BU5, 00000512868BU1, 00000512870BU7, 00000512871BU5, 00000512874BU9, 00000512875BU6, 00000512877BU2, 00000512880BU6, 00000512883BU0, 00000512888BU9, 00000512891BU3, 00000512892BU1, 00000512894BU7, 00000512898BU8, 00000513110BU7, 00000513118BU0, 00000513161BU0, 00000513190BU9, 00000513204BU8, 00000513206BU3, 00000513207BU1, 00000513208BU9, 00000513209BU7, 00000513212BU1, 00000513244BU4, 00000513265BU9, 00000513274BU1, 00000513296BU4, 00000513516BU5, 00000513533BU0, 00000513539BU7, 00000513541BU3, 00000513547BU0, 00000513555BU3, 00000513558BU7, 00000513560BU3, 00000513567BU8, 00000519094BU7, 00000519100BU2, 00000519106BU9, 00000519109BU3, 00000519121BU8, 00000519125BU9, 00000519129BU1, 00000519131BU7, 00000520045BU6, 00000520048BU0, 00000520050BU6, 00000520065BU4, 00000520070BU4, 00000520075BU3, 00000520083BU7, 00000526275BU3, 00000526389BU2, 00000526556BU6, 00000526557BU4, 00000526560BU8, 00000526562BU4, 00000526563BU2, 00000526648BU1, 00000526681BU2, 00000529461BU6, 00000530106BU4, 00000530116BU3, 00000530120BU5, 00000530242BU7, 00000530246BU8, 00000530248BU4, 00000530410BU0, 00000530413BU4, 00000536802BU2, 00000538053BU0, 00000543790BU0, 00000544289BU2, 00000545052BU3, 00000546574BU5, 00000547017BU4. systems id: 835160017, 82416070006, 00192VAS05, 214947MINNOVA2, 305682IN2X, 318448INNOVA3, 83INNOVA2588441, B5219603, H0002VAS66, HU1099VA02, RPINNOVA2100, YV0053.
510(k) numbers
["K050489","K052157","K052412"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2183066
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register