Skip to the content

FDA Medical Device Recalls (openFDA)

Avid Medical, Inc.: Custom procedure trays containing medical components needed to facilitate patient procedures and to perform minor and major patient surgeries. Multiple item and lot numbers.

Recall number
Z-1516-2020
Recalling firm
Avid Medical, Inc.
Product
Custom procedure trays containing medical components needed to facilitate patient procedures and to perform minor and major patient surgeries. Multiple item and lot numbers.
Status
Terminated
Initiated
2020-02-18
Posted
not on file
Terminated
2020-11-23
Reason
The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.
Root cause
Other
Action
URGENT MEDICAL DEVICE RECALL NOTICE letters, dated February 18, 2020, were distributed to customers. Letter provides identification of the issue, health risk and action to take. Customers were instructed to take the following required actions: " Review all storage and usage locations and segregate and quarantine any trays with an item code and lot number listed on attached Exhibit A. " Complete and return the enclosed Recall Response Form to Sharon.moll@owens-minor .com. Please respond even if you do not have any trays subject to this recall. " Upon receipt of your completed Recall Response Form, product return instructions will be provided and a RGA will be issued, if you do have trays subject to this recall. " If you have further distributed this product, please identify your customers and notify them at once of this product recall. Your notification to your customers may be enhanced by including a copy of this Urgent Medical Device Recall Notice . Your assistance is appreciated and necessary to prevent any unintended use of the product. If you have any questions, call Sharon Moll at 757-566-9844.
Quantity
1,906 each
Distribution
US: AZ, CA, IL, IN, MD, MS, NJ, NV, NY, TN, TX WI, VA
Lot, serial or model codes
Item Codes:ADVC033-16, ADVG034-15, ADVU018-24, JNTH001-04, JNTH001-05, JNTH002-05, SHER019-07, YAVA100-05. Lot Numbers: 1344066, 1344358, 1353858, 1353860, 1354012, 1359439, 1361116, 1361767, 1363232, 1365746, 1379393. Expiration dates: 2/28/21 thru 7/30/2022.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OES
Firm FEI
1047429
Firm city
Toano
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register