Siemens Healthcare Diagnostics, Inc.: Stratus(R) CS Acute care(TM) NT - proBNP TestPak reagent (CPBNPM) The Stratus(R) CS Acute care(TM) NT - proBNP method (pBNP) is an in vitro diagnostic test for the quantitative measurement of N-terminal pro-brain natriuretic peptide (NT0proBNP) in heparinized plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patient with acute coronary syndrome and heart failure.
- Recall number
- Z-1516-2012
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Stratus(R) CS Acute care(TM) NT - proBNP TestPak reagent (CPBNPM) The Stratus(R) CS Acute care(TM) NT - proBNP method (pBNP) is an in vitro diagnostic test for the quantitative measurement of N-terminal pro-brain natriuretic peptide (NT0proBNP) in heparinized plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patient with acute coronary syndrome and heart failure.
- Status
- Terminated
- Initiated
- 2012-04-05
- Posted
- 2012-05-10
- Terminated
- 2013-03-19
- Reason
- Firm has confirmed a low frequency of calibration failures due to slope errors that are not resolved with routine maintenance with the specific SCS Pro-BNP lots listed. This issue only causes the slope error failures but does not affect patient results when the calibration is within specifications.
- Root cause
- Component design/selection
- Action
- The firm, SIEMENS, sent a "Urgent Field Safety Notice" dated March, 2012 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to discontinue use of the lots once revised expiration date occurs; contact their local Siemens Customers Service Center at 888-588-3916 for replacement; complete and return the FIELD CORRECTION EFFECTIVENESS CHECK form via fax to the TECHNICAL SOLUTIONS CENTER at (302) 631-8467 as soon as possible, and forward this notification to anyone to whom they may have distributed this product. The remaining inventory at the firm was destroyed. If you have any technical questions regarding this information, please contact the Siemens Technical Solutions Center at 800-405-6473.
- Quantity
- 1,899 cartons
- Distribution
- Nationwide distribution: USA including states of: AK, AR, CA, CO, CT, DC, FL, HI, ID, IN, IL, IA, KS, KY, LA, ME, MI, MN, MS, MO, MT, NE, NH, NM, ND, NY, OH, OK, OR, PA, SD, TX, UT, VT, WA, WV, WI, and WY.
- Lot, serial or model codes
- Lot numbers 211214002, exp 04/28/2012; 211227002, exp 05/11/2012; 211283002, 07/06/2012; 211297002, 07/20/2012; 211339002, 08/31/2012; 212003002, exp 09/29/2012; and 212023002, exp 10/19/2012.
- 510(k) numbers
- ["K071834"]
- PMA numbers
- not on file
- Product code
- NBC
- Firm FEI
- 2517506
- Firm city
- Newark
- Firm state
- DE
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28