Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare: GE Healthcare Innova 3100 Digital Fluoroscopic Imaging System with Bolus Chasing option.

Recall number
Z-1516-2008
Recalling firm
GE Healthcare
Product
GE Healthcare Innova 3100 Digital Fluoroscopic Imaging System with Bolus Chasing option.
Status
Terminated
Initiated
2007-10-21
Posted
2008-09-29
Terminated
2011-12-17
Reason
Incorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option) may be underestimated by up to 50%. In order to avoid this issue, it is necessary to perform a system reset or reboot every day. GE Healthcare is working on a c
Root cause
Nonconforming Material/Component
Action
GE sent all customers with affected units a letter on 11/15/07 titled Urgent Medical Device Correction. In the US this was sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX overnight Delivery). The letter was sent to 3 titles within the affected accounts; Hospital Administrator, Manager of Radiology/Cardiology and Radiologist/ Cardiologist. A correction letter identified a short term recommendation to perform a system reset or reboot every day and stated that a software correction will be installed at their site at no cost.
Quantity
146 total
Distribution
Worldwide Distribution: USA, Argentina, Australia, Canada, China, Germany, Japan, Mexico and Russia.
Lot, serial or model codes
8613100, 77573841, 0910064009, 3018774100, 4103323100, 4103374100, 4103623100, 4105434100, 4438494100, 5138534100, 5703214100, 7083333100, 7278253100, 8019643100, 8595724100, 9408984100, 082416040017, 082416070004, 082416120006, 082416130013, 082416140003, 201541LAB2, 229228INNOV41V, 239436CATH1, 251435ICV5, 270688INN, 304255INNOVA, 304347INNOVA, 309691P3100, 310423INNOVA1, 313593IC1, 313593IC2, 314525SACL2, 315448RM4, 315448RM5, 316268INNOVAA, 319338VA4100, 325670CATH, 334286BSP, 352333VASC, 352376SP, 352401CATH, 352401VASC, 378INV31, 407SPECIALS, 409899LAB3, 413447CATH, 414961INN4100, 416756INNOVA, 417269SCLD, 478633INOV, 502587LAB4, 508856OR41, 516562INNOVA4, 516663INNOVA2, 518525LAB1, 540953INNOVA, 561266IN31, 563421CATHLAB8, 573632CRA, 573761SMA, 573815BCR1, 574535IN41, 574647IN41, 6012884100SP1, 601553CL3, 602239LAB4, 602239LAB6, 603778ECL1, 606833INNOVA, 614293LIBERTY1, 614293LIBERTY2, 615338CV2, 615769SKYCV2, 619340VA8, 619482INNOVA1, 619541INNOVA31, 619552INNOVA1, 626359INNOVA, 702616INNOVA, 7028803100A, 706774CCL5, 713794LAB7, 713794LAB9, 715369INN4100, 717217INN31, 717763INNOVA41, 717763OR4100, 718283INNOVA3, 718518CVL1, 724656ANG, 724773C1, 727725SPEC1, 731422CV1, 732294INOV, 757398INNOVA, 757736INNOVA, 765453INNOVA1, 770991CL3, 773792CL3100, 773NORWG3100, 803329CL31, 803329CL41, 804594CATH, 808547QCV2A, 812238LIB1, 812279INNOVA1, 812283IN31, 812450EP1, 812482INNOVA, 813972INTV, 816276SP1, 817255NH4100, 817433HSW3100, 83INNOVA1188379, 845365V19, 850678IN3100, 865541CCL5, 865693CCL3, 901726SP2, 903CSM4100, 904202CATH3, 904276SP, 908522LAB1, 909788INNOVA1, 918307SHIN4, 918744JC4100, 928779INNOVA1, 936266CL1, 956389INN31, 956632INN4100, 956994INN1, 956994INN2, 972487ANGIO1, 973971LAB3, BPINNOVA2, BPINNOVA3R, BPINNOVA3R, ECXV1406C, FHORLCL1, HC1536XR04, RU1550VA01, RU2457VA01, SLIN4100, YV0022, and YV0036.
510(k) numbers
["K042053"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register