Roche Diagnostics Operations, Inc.: cobas e801 Immunoassay Analyzer

Recall number
Z-1515-2022
Recalling firm
Roche Diagnostics Operations, Inc.
Product
cobas e801 Immunoassay Analyzer
Status
Open, Classified
Initiated
2022-07-14
Posted
not on file
Terminated
not on file
Reason
Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo IgG.
Root cause
Software design
Action
The firm distributed an Urgent Medical Device Correction notification via mail by UPS on 07/14/2022. The notice indicates that cobas e801 (Catalog No. 08454345001) and cobas e801 module (Catalog No. 04682913001) are being recalled due to a software issue affecting all analyzers, which may result in impacted signals and sample results of Pre-Wash tests when using Quick Start Mode for the following assays: Anti-HAV 2, Anti-HBc IgM, IGF-1, Myoglobin, Rubella IgG, and Toxo IgG. Customers with affected units are provided workaround instructions to deactivate Quick Start Mode or additional software maintenance with Quick Start Mode still activated. The firm states that the issue will be fixed in the next software version release. Customers are asked to return the attached Faxback form. Any questions are to be directed to Roche Customer Service at 1-800-428-2336.
Quantity
965 units
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, DC & PR.
Lot, serial or model codes
UDI-DI: 07613336158456 and 04015630946198; Catalog No. 08454345001 and 07682913001. All serial numbers.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJE
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register