Skip to the content

FDA Medical Device Recalls (openFDA)

CSA Medical: CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device Model: CC3-01

Recall number
Z-1513-2015
Recalling firm
CSA Medical
Product
CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device Model: CC3-01
Status
Terminated
Initiated
2015-03-25
Posted
2015-04-23
Terminated
2015-08-21
Reason
TruFreeze Console caused a higher rate of liquid nitrogen (cryogen) to be delivered and may cause: stricture,scarring, bradycardia, or pneumothorax
Root cause
Software design
Action
CSA Medical issued letter dated 3/25/15 advising users of the problem. Users provided with: The mitigations available to the active venting procedures coupled with the extremely unlikely probability of injury, rare risk, thus allowing the physician to continue with active venting procedures. In regards a passive venting procedure may not have sufficient mitigation to allow the physician to identify the potential hazard with sufficient time to preclude a potential for injury. Therefore, passive users will be instructed not to use the system until a software improvement is put into place. Additionally, no catheters or consoles will be shipped to passive venting users of truFreeze. A response form to be signed and returned confirming receipt of the notification.Questions contact: Stephen Mascioli, MD 781-538-4755 smascioli@csamedical.com.
Quantity
82 units
Distribution
Nationwide
Lot, serial or model codes
Serial Numbers:  01-00106 through 01-00201
510(k) numbers
["K143625"]
PMA numbers
not on file
Product code
GEH
Firm FEI
3010140265
Firm city
Lexington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register