Brainlab AG: ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED 49973B ET DATA PREP / REVIEW SYSTEM 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED
- Recall number
- Z-1513-2014
- Recalling firm
- Brainlab AG
- Product
- ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED 49973B ET DATA PREP / REVIEW SYSTEM 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED
- Status
- Terminated
- Initiated
- 2014-02-24
- Posted
- 2014-04-25
- Terminated
- 2017-04-24
- Reason
- When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.
- Root cause
- Software design
- Action
- A FIELD SAFETY NOTICE / PRODUCT INFORMATION letters, dated February 24, 2014, were sent to all direct accounts. The letter identified the affected product, the reason for the recall, and included instructions for direct accounts to take while a permanent fix can be deployed. The letter also advised that a software update will be made available starting the end of April 2014. Customers with questions or concerns can contact Brainlab at 800-597-5911 or via e-mail at us.support@brainlab.com.
- Quantity
- 41 systems (13 systems in US; 28 systems to foreign countries)
- Distribution
- Worldwide Distribution -- USA, including the states of CA, IN, LA, MA, OH, TX, and WI, and the countries of Australia, China, France, Germany, Japan, Netherlands, and the United Kingdom.
- Lot, serial or model codes
- ExacTrac version v.6.0.3.
- 510(k) numbers
- ["K120789"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 3002619595
- Firm city
- Feldkirchen
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28