Qapel Medical Inc.: Hippo 072 Aspiration System including Cheetah Delivery Tool and Aspiration Tubing, REF: APT6072-132. Hippo 072 Aspiration System including Cheetah Delivery Tool, REF: AP6072-132. 072 Aspiration Tubing, REF: APT-95

Recall number
Z-1512-2025
Recalling firm
Qapel Medical Inc.
Product
Hippo 072 Aspiration System including Cheetah Delivery Tool and Aspiration Tubing, REF: APT6072-132. Hippo 072 Aspiration System including Cheetah Delivery Tool, REF: AP6072-132. 072 Aspiration Tubing, REF: APT-95
Status
Open, Classified
Initiated
2025-02-26
Posted
2025-04-07
Terminated
not on file
Reason
Aspiration catheter distal tip features and characteristics may not be in the scope of FDA clearance; tip detachment, vessel rupture, and vasospasm could occur.
Root cause
Device Design
Action
On 2/26/2025, removal and discontinuation notices were mailed to customers were asked to do the following: 1) Quarantine affected devices and return to the firm. 2) If you have provided this product to others, let the firm know so they can notify them of this action and retrieve any remaining product. 3) Complete and return the Acknowledgement and Response Card via email to orders@qapelmedical.com If you have any questions regarding this letter, please contact customer service for further information at orders@qapelmedical.com, or by phone at 510-738-6255 between the hours of 8am - 5pm PST.
Quantity
1617
Distribution
US: CA, IL, NE, TN, TX, NJ, NC, SD, FL, MI, PA, GA, AL, IN, NY, WI, AZ, MN, LA, OH, OK, SC, NV, ND, ID, MO, CT, WA, MA. OUS: QATAR, United Arab Emirates, Republic of Kazakhstan
Lot, serial or model codes
REF/UDI-DI/Lot(Expiration): APT6072-132/00857545008127/FG241008C-03(04/07/2025), FG240916C-04(03/17/2025), FG240905C-04(03/06/2025). AP6072-132/00857545008097/FG241206A-03(06/08/2025), FG240917A-01(03/17/2025). APT-95/00857545008103/FG241206A-04(06/08/25)
510(k) numbers
["K222786"]
PMA numbers
not on file
Product code
NRY
Firm FEI
3020954126
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register