Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Boston Scientific, Implantable Cardioverter Defibrillator, DYNAGEN EL ICD DR, Model D152 - Product Usage: Implantable cardioverter defibrillators (ICDs) are intended to provide ventricular anti-tachycardia pacing (ATP) and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.

Recall number
Z-1512-2020
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific, Implantable Cardioverter Defibrillator, DYNAGEN EL ICD DR, Model D152 - Product Usage: Implantable cardioverter defibrillators (ICDs) are intended to provide ventricular anti-tachycardia pacing (ATP) and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
Status
Terminated
Initiated
2020-02-12
Posted
2020-03-03
Terminated
2023-04-14
Reason
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved by the US Food and Drug Administration.
Root cause
Under Investigation by firm
Action
The firm initiated the recall by letter on 12/12/2020. The letter explained the error and requested the consignee segregate the units for removal. The firm will provide another model of programmer to be used in the interim.
Quantity
1 device
Distribution
US Nationwide distribution in the states of NC, TN, and MO.
Lot, serial or model codes
Serial Number 516959
510(k) numbers
not on file
PMA numbers
["P960040S306"]
Product code
LWS
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register