FDA Medical Device Recalls (openFDA)
Boston Scientific Corporation: Boston Scientific, Implantable Cardioverter Defibrillator, DYNAGEN EL ICD DR, Model D152 - Product Usage: Implantable cardioverter defibrillators (ICDs) are intended to provide ventricular anti-tachycardia pacing (ATP) and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
- Recall number
- Z-1512-2020
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific, Implantable Cardioverter Defibrillator, DYNAGEN EL ICD DR, Model D152 - Product Usage: Implantable cardioverter defibrillators (ICDs) are intended to provide ventricular anti-tachycardia pacing (ATP) and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
- Status
- Terminated
- Initiated
- 2020-02-12
- Posted
- 2020-03-03
- Terminated
- 2023-04-14
- Reason
- Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved by the US Food and Drug Administration.
- Root cause
- Under Investigation by firm
- Action
- The firm initiated the recall by letter on 12/12/2020. The letter explained the error and requested the consignee segregate the units for removal. The firm will provide another model of programmer to be used in the interim.
- Quantity
- 1 device
- Distribution
- US Nationwide distribution in the states of NC, TN, and MO.
- Lot, serial or model codes
- Serial Number 516959
- 510(k) numbers
- not on file
- PMA numbers
- ["P960040S306"]
- Product code
- LWS
- Firm FEI
- 2124215
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as