Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Irix Gamma Camera, Scintillation (gamma) camera, Philips Medical System, Cleveland, OH 44143

Recall number
Z-1511-2008
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Irix Gamma Camera, Scintillation (gamma) camera, Philips Medical System, Cleveland, OH 44143
Status
Terminated
Initiated
2007-11-21
Posted
2008-09-19
Terminated
2012-02-25
Reason
Resistor Failure: The Rotate Motion Shunt Resistor may overheat and results in the appearance of smoke, a burning smell, and an electronic stop condition, which will disable the gamma camera rotate motion and all other motions.
Root cause
Component design/selection
Action
Consignees were sent an Urgent Field Safety Notice on 1/30/08. The letter instructs users who experience a failure to remove the system from service and contact the recalling firm for immediate correction. The recalling firm will schedule appointments with users to replace the affected part beginning in March 2008. For additional information, contact 1-800-722-9377.
Quantity
Unknown at this time sinc the recall was expanded.
Distribution
Worldwide Distribution --- USA including states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV, Puerto Rico and countries of Argentinia, Austrailia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iceland, India, Ireland, Israel, Italy, Japan, Netherlands, Norway, Philipines, Poland, Singapore, South Africa, Spain, Switzerland, and United Kingdom.
Lot, serial or model codes
Model Numbers: 210857 and 210881.  4535 679 46991, or N210857 Irix System, 3/8" Crystal 4535 665 13551, or N210881 Irix System 3/4" Crystals 4535 679 46981, or N211038 Irix System 3/8 Precision 4535 665 13571, or N211040 Irix System 3/4 Precision (Note: Due to a part number scheme change at the firm, either number is representative of the identified system)
510(k) numbers
["K964712"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register