Skip to the content

FDA Medical Device Recalls (openFDA)

OraSure Technologies, Inc.: OraQuick HCV Visual Reference Panel Intended to assist new operation in becoming proficient at reading specimens with antibody levels near the limit of detection of the device.

Recall number
Z-1509-2016
Recalling firm
OraSure Technologies, Inc.
Product
OraQuick HCV Visual Reference Panel Intended to assist new operation in becoming proficient at reading specimens with antibody levels near the limit of detection of the device.
Status
Terminated
Initiated
2016-01-15
Posted
not on file
Terminated
2018-02-07
Reason
OraSure Technologies, Inc. discovered the package insert included with the OraQuick HCV Rapid Antibody test Visual Reference Panel may be incorrect due to a complaint received from one of their customers. The customer reported they received the Ora Quick Ebola Visual Reference Panel instead of an OraQuick HCV Visual Reference Panel.
Root cause
Unknown/Undetermined by firm
Action
Letters, dated January 15, 2016, and correct HCV VRP insert were sent to all customers who received HCV VRP lot 6648737. The letter stated the reason for the recall; and, asked users to discard the Package Insert that was included with the product and replace it with the corrected one provided with the letter. Questions regarding the recall should be directed to OraSure Sales Representative or our Customer Care Department at 1-800-ORASURE.
Quantity
244 Visual Reference Panel Kits
Distribution
Nationwide Distribution -- AK, AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, LA, MA, MI, MO, MT, NC, ND, NE, NH, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, VA, WA, and WI.
Lot, serial or model codes
Item Number 1001-0343, Lot # 6648737
510(k) numbers
not on file
PMA numbers
["P080027"]
Product code
MZO
Firm FEI
2528909
Firm city
Bethlehem
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register