Cordis Corporation: FLEXSTENT Biliary Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Biliary SES product is indicated for use in the palliation of malignant strictures in the biliary tree.

Recall number
Z-1509-2013
Recalling firm
Cordis Corporation
Product
FLEXSTENT Biliary Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Biliary SES product is indicated for use in the palliation of malignant strictures in the biliary tree.
Status
Terminated
Initiated
2013-04-23
Posted
2013-06-06
Terminated
2014-06-27
Reason
Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.
Root cause
Package design/selection
Action
Cordis sent an Urgent Medical Device Correction and Potential Removal letter dated April 24, 2013 e-mail to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately identify and set aside all affected products to prevent further use of the affected product, until inspected. Use only product identified with a blue dot next to the lot number on the carton face label, carton end label, and outer pouch label, or product with upgraded sterile barrier system (not yet available, as of the date of this letter), review, complete, sign and return the enclosed Acknowledgement Form and maintain a copy of this notice with the affected product. For questions related to the issue, please contact our FSS/Cordis representative, or please contact Cordis Direct at (800) 781-0282, Monday through Friday from 9:00 AM to 5:00 PM EDT.
Quantity
1357 units
Distribution
Worldwide Distribution - USA (nationwide) and the states of: AZ, CA, CT, FL, GA, IA, IL, LA, MD, MI, MS, NC, NJ, NY, OH, PA, RI, SC, TX, WA, WI, WV, and the countries of: Czech Republic, Italy, Netherlands, and New Zealand
Lot, serial or model codes
27 Catalog Numbers   FLX-50030-B2, FLX-60030-B2, FLX-70030-B2, FLX-80030-B2, FLX-50040-B2, FLX-60040-B2, FLX-70040-B2, FLX-80040-B2, FLX-50060-B2, FLX-60060-B2, FLX-70060-B2, FLX-80060-B2, FLX-50080-B2, FLX-60080-B2, FLX-70080-B2, FLX-80080-B2, FLX-50100-B2, FLX-60100-B2, FLX-70100-B2, FLX-80100-B2, FLX-50120-B2, FLX-60120-B2, FLX-70120-B2, FLX-80120-B2, FLX-60150-B2, FLX-70150-B2, FLX-80150-B2
510(k) numbers
["K112522"]
PMA numbers
not on file
Product code
FGE
Firm FEI
1016427
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register