Skip to the content

FDA Medical Device Recalls (openFDA)

Guangdong Haiou Medical Apparatus Co., Ltd: HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25

Recall number
Z-1506-2022
Recalling firm
Guangdong Haiou Medical Apparatus Co., Ltd
Product
HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25
Status
Open, Classified
Initiated
2022-06-10
Posted
not on file
Terminated
not on file
Reason
Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user
Root cause
No Marketing Application
Action
Guangdong Haiou Medical., LTD (Haiou) contacted direct consignee by email on [10-Jun-2022] and Formal Recall Urgent Medical Device Recall letter through email and first-class letter. Letter states reason for recall, health risk and action to take: " Remove product from sale and cease distribution of product subject to this recall action. " Immediately review your inventory for the specific Product Description and Lot Number listed above and place into quarantine until further notice. " Notify us of any stock you have still under your direct control " Notify all customers to whom you have distributed this product of this recall. Use the template provided with this letter. " Complete the attached Customer Response Form and return to the Haiou contact noted on the form whether or not you have any of the impacted material so that Haiou may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement. " Forward any customer complaints for the use of this product to Haiou as per the contact information provided. Jiehong Luo QA Manager Phone:+ 86-663-2900999 Email: qc@haiou.net.cn Address: Nanyuan Industrial Zone, North Liusha, Puning City, Guangdong, China
Quantity
4,050,000 units
Distribution
US Nationwide distribution in the state of CA.
Lot, serial or model codes
Lot Number: 20IB0 Exp. Date: 2025/09/04
510(k) numbers
not on file
PMA numbers
not on file
Product code
MEG
Firm FEI
3010596027
Firm city
Puning
Firm state
not on file
Firm country
China

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register