Skip to the content

FDA Medical Device Recalls (openFDA)

Alcon Grieshaber AG: The Constellation Vision System Pneumatic Handpiece. Indicated for both anterior and posterior segment opthalmic surgery.

Recall number
Z-1506-2013
Recalling firm
Alcon Grieshaber AG
Product
The Constellation Vision System Pneumatic Handpiece. Indicated for both anterior and posterior segment opthalmic surgery.
Status
Terminated
Initiated
2013-04-17
Posted
2013-06-06
Terminated
2015-01-07
Reason
Alcon is conducting a Medical Device Removal for the Constellation Pneumatic Handpiece, DSP, product number 725.01. This action is being taken because the forceps or scissors tips (sold separately) may dislodge from the hand piece, if not attached by the user according to the Directions for Use.
Root cause
Use error
Action
Alcon sent an "URGENT: MEDICAL DEVICE REMOVAL" letter on April 17, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact your local Alcon Sales Representative for questions regarding this notice.
Quantity
8,455 units
Distribution
Worldwide distribution: US (nationwide) and countries of: Argentina, Australia, Austria, Belgium, Canada, Chile, China, Egypt, France, Germany, Great Britain, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Kyrgyzstan, Lebanon, Korea, Malaysia, Mexico, Pakistan, Peru, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Turkistan, Venezuela, and Veit Nam.
Lot, serial or model codes
All Lots of the Constellation Vision System Pneumatic Handpiece, DSP, product number 725.01
510(k) numbers
not on file
PMA numbers
not on file
Product code
HQE
Firm FEI
3003398873
Firm city
Schaffhausen
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register