FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: MULTIX FUSION. The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.
- Recall number
- Z-1505-2014
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- MULTIX FUSION. The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.
- Status
- Terminated
- Initiated
- 2014-02-13
- Posted
- 2014-04-24
- Terminated
- 2015-03-16
- Reason
- It was observed that the corrugated hose holder above the 3D rail on the Multix Fusion system became loose. According to the Installation Instructions the corrugated hose holder shall be afixed with transversal carriage, with Loctite 243 and Torque 10nm during system installation. Siemens initiated the update instruction to verify and secure the hose holder with Loctite 243. A PCB D802 Rev. 00 b
- Root cause
- Process change control
- Action
- Affected customers were initially notified with a recall letter on/about 2/12/14 and were sent Update Instructions to affected customers.
- Quantity
- 12
- Distribution
- Distributed in the states of AK, PA, KY, OH, WI, MI, FL, and NY.
- Lot, serial or model codes
- model nos. 10746665, 10746666, 10847610, 10746700
- 510(k) numbers
- ["K121513"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28