Siemens Medical Solutions USA, Inc: MULTIX FUSION. The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.

Recall number
Z-1505-2014
Recalling firm
Siemens Medical Solutions USA, Inc
Product
MULTIX FUSION. The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.
Status
Terminated
Initiated
2014-02-13
Posted
2014-04-24
Terminated
2015-03-16
Reason
It was observed that the corrugated hose holder above the 3D rail on the Multix Fusion system became loose. According to the Installation Instructions the corrugated hose holder shall be afixed with transversal carriage, with Loctite 243 and Torque 10nm during system installation. Siemens initiated the update instruction to verify and secure the hose holder with Loctite 243. A PCB D802 Rev. 00 b
Root cause
Process change control
Action
Affected customers were initially notified with a recall letter on/about 2/12/14 and were sent Update Instructions to affected customers.
Quantity
12
Distribution
Distributed in the states of AK, PA, KY, OH, WI, MI, FL, and NY.
Lot, serial or model codes
model nos. 10746665, 10746666, 10847610, 10746700
510(k) numbers
["K121513"]
PMA numbers
not on file
Product code
KPR
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register