FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: BD BBL Staphyloslide Latex Test Kit, Catalog # 240952 (100 tests/kit) and Catalog # 240953 (500 tests/kit), packaged in a kit configuration with test reagents and labware, Becton, Dickinson and Company, 7 Loveton Circle Sparks, Maryland 21152, USA , 800-638-8663, BENEX Limited, Shannon, County Clare, Ireland.

Recall number
Z-1505-06
Recalling firm
Becton Dickinson & Co.
Product
BD BBL Staphyloslide Latex Test Kit, Catalog # 240952 (100 tests/kit) and Catalog # 240953 (500 tests/kit), packaged in a kit configuration with test reagents and labware, Becton, Dickinson and Company, 7 Loveton Circle Sparks, Maryland 21152, USA , 800-638-8663, BENEX Limited, Shannon, County Clare, Ireland.
Status
Terminated
Initiated
2006-08-17
Posted
2006-09-16
Terminated
2007-03-09
Reason
False Negative Staphylococcus aureus results which could prevent infected patients from receiving antimicrobial treatment; due to the failure of the Positive Control to react within 20 seconds. The failure is due to reduced reactivity of the latex component.
Root cause
Other
Action
The recalling firm notified distributors and end users by e-mail, faxed letters, and certified courier on 08/17/06.
Quantity
1,572 kits
Distribution
Worldwide-product was sold to distributors and end users
Lot, serial or model codes
Lot Number / Exp. Date 16629910 April 30, 2007-Cat # 240952 16562410 April 30, 2007-Cat # 240952 16629920 April 30, 2007-Cat # 240953 16562420 April 30, 2007-Cat # 240953
510(k) numbers
not on file
PMA numbers
not on file
Product code
LHJ
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register