Philips Respironics, Inc.: Trilogy EV300, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.

Recall number
Z-1502-2024
Recalling firm
Philips Respironics, Inc.
Product
Trilogy EV300, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
Status
Open, Classified
Initiated
2024-03-19
Posted
2024-04-17
Terminated
not on file
Reason
Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Root cause
Device Design
Action
Consignees were mailed an URGENT MEDICAL DEVICE CORRECTION notice, dated 3/7/24. Consignees are asked to identify any affected product in their inventory and follow the provided instructions in order to continue using devices safely in CPAP or PSV mode. Devices that alarm are to be immediately plugged into a power source if a Loss of Power Alarm occurs. If no source of power is available, remove and replace the detachable battery. Devices running on software other than 1.05.06.00 can be updated to the latest software version, which does not experience this failure mode. The provided recall notice should be shared with all members of consignee organizations that utilize the recalled devices. If product was further distributed the provided notice should be forwarded. Consignees with any questions can contact Philips Respironics Customer Service at 1-800-345-6443 for homecare customers or 1-800-722-9377 for hospital customers.
Quantity
56,012 units
Distribution
Domestic: Nationwide Distribution; Foreign: AE, AR, AT, AU, BD, BE, BG, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, US, UY, VN, YT, ZA, ZW.
Lot, serial or model codes
Model No. BL2200X15B, BR2200X18B, CA2200X12B, EE2200X15B, EU2100X15B, EU2100X19, EU2200X15B, FR2200X14B, DE2200X13B, GB2200X15B, FX2100X15B, FX2200X15B, IN2100X15B, IN2100X19, IN2200X15B, IT2200X21B, JP2100X16B, KR2200X15B, LA2100X15B, ND2200X15B, ES2200X15B, TR2200X15B, DS2100X11B, DS2200X11B, UDI-DI: 606959058668, 606959058583, 606959052321, 606959058705, 606959055674, 606959055681, 606959058682, 606959058613, 606959058651, 606959058675, 606959061019, 606959061033, 606959054059, 606959055575, 606959056497, 606959058637, 606959055582, 606959058590, 606959055599, 606959058620, 606959058644, 606959058798, 606959051997, 606959052017; All units except those with software version 1.05.06.00.
510(k) numbers
["K181166"]
PMA numbers
not on file
Product code
CBK
Firm FEI
2518422
Firm city
Murrysville
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register