FDA Medical Device Recalls (openFDA)
Philips Respironics, Inc.: Trilogy EV300, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
- Recall number
- Z-1502-2024
- Recalling firm
- Philips Respironics, Inc.
- Product
- Trilogy EV300, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
- Status
- Open, Classified
- Initiated
- 2024-03-19
- Posted
- 2024-04-17
- Terminated
- not on file
- Reason
- Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
- Root cause
- Device Design
- Action
- Consignees were mailed an URGENT MEDICAL DEVICE CORRECTION notice, dated 3/7/24. Consignees are asked to identify any affected product in their inventory and follow the provided instructions in order to continue using devices safely in CPAP or PSV mode. Devices that alarm are to be immediately plugged into a power source if a Loss of Power Alarm occurs. If no source of power is available, remove and replace the detachable battery. Devices running on software other than 1.05.06.00 can be updated to the latest software version, which does not experience this failure mode. The provided recall notice should be shared with all members of consignee organizations that utilize the recalled devices. If product was further distributed the provided notice should be forwarded. Consignees with any questions can contact Philips Respironics Customer Service at 1-800-345-6443 for homecare customers or 1-800-722-9377 for hospital customers.
- Quantity
- 56,012 units
- Distribution
- Domestic: Nationwide Distribution; Foreign: AE, AR, AT, AU, BD, BE, BG, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, US, UY, VN, YT, ZA, ZW.
- Lot, serial or model codes
- Model No. BL2200X15B, BR2200X18B, CA2200X12B, EE2200X15B, EU2100X15B, EU2100X19, EU2200X15B, FR2200X14B, DE2200X13B, GB2200X15B, FX2100X15B, FX2200X15B, IN2100X15B, IN2100X19, IN2200X15B, IT2200X21B, JP2100X16B, KR2200X15B, LA2100X15B, ND2200X15B, ES2200X15B, TR2200X15B, DS2100X11B, DS2200X11B, UDI-DI: 606959058668, 606959058583, 606959052321, 606959058705, 606959055674, 606959055681, 606959058682, 606959058613, 606959058651, 606959058675, 606959061019, 606959061033, 606959054059, 606959055575, 606959056497, 606959058637, 606959055582, 606959058590, 606959055599, 606959058620, 606959058644, 606959058798, 606959051997, 606959052017; All units except those with software version 1.05.06.00.
- 510(k) numbers
- ["K181166"]
- PMA numbers
- not on file
- Product code
- CBK
- Firm FEI
- 2518422
- Firm city
- Murrysville
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28