Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Ysio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.

Recall number
Z-1501-2020
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Ysio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.
Status
Terminated
Initiated
2020-02-10
Posted
2020-03-03
Terminated
2020-08-31
Reason
The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the systems controls may (depending on the values being processed) initiate a correction of the tube stand position, resulting in an unexpected movement.
Root cause
Component design/selection
Action
On February 10, 2020, the firm notified affected customers via Urgent Medical Device Correction letters. Customers were informed of the product issue, and instructed to take the following actions: When using the system, users should follow the regular cautions and instructions listed in the system operator manual, particularly in chapter Safety, section Information about unit movements. In addition, users and all accompanying personnel are to pay special attention to remain outside of the zone of potential device movement (hazardous zone), e.g., remain outside the possible movement area of longitudinal and transversal direction of the ceiling stand. Siemens Healthineers has issued a field software update that will eliminate the issue described. The resolution will be provided free of charge. The software update is already available and is being distributed. Your local Siemens Healthineers Service organization will contact you to schedule the software update for your system. To schedule an earlier appointment, you may contact the firm's service organization at 1-800-888-7436.
Quantity
81 US/ 308 worldwide
Distribution
US Nationwide distribution in the states of AR, CA, DC, FL, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, UT, VA and WI.
Lot, serial or model codes
Serial Nos. 20113 20034 21100 21086 21113 21088 21103 21221 21223 21224 21085 20045 20093 20110 21066 21031 21139 21006 20155 21112 21122 21185 20088 20089 20106 20107 21052 21089 21213 20167 21098 21174 21057 21101 21116 21117 21118 21120 21091 21092 21037 21076 21077 21200 21203 21209 20149 21163 21172 21171 21173 20094 21115 20047 21194 20193 21211 21212 21186 21188 21094 21039 21040 21108 20035 20185 21061 21065 21205 21095 21148 21189 21140 21175 21023 21001 21127 21123 21013 20159 21160
510(k) numbers
["K081722"]
PMA numbers
not on file
Product code
KPR
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register