Haag-Streit USA Inc: EyeSuite i.8.2.1.0 Software for ophthalmic use including selection of Intra Ocular Lenses (IOLs).

Recall number
Z-1500-2017
Recalling firm
Haag-Streit USA Inc
Product
EyeSuite i.8.2.1.0 Software for ophthalmic use including selection of Intra Ocular Lenses (IOLs).
Status
Terminated
Initiated
2017-01-05
Posted
not on file
Terminated
2019-03-22
Reason
There is a possibility for data to be stored under the wrong patient on the DICOM Server following a non-standard workflow.
Root cause
Software design
Action
On January 5, 2017 the firm sent Field Safety Notice to their customers.
Quantity
163 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Model # 7220374 EyeSuite DICOM Option, 7220379 DICOM License .BQ 900 LED S/N: 3682  BQ LET Table S/N: 10038  BX 900 LED w/Digital EyeSuite S/N: 10016  CM 900 S/N: 354 & 491.  IM 900 / BQ 900 LED S/N: 919, 5036, 1136, 3146, 3147, 3148, 3150, 560, 3131, 2412 & 3011.  License Only S/N: 03011, 3210, BP5P922, 5617, 724, 1406, 3204, 2306, 2314, 2514, 3087, 1246, 2395 & 1063.  Lenstar S/N: 167, 285, 332, 416, 469, 656, 878, 1134, 1136, 1197, 1313, 1332, 1607, 2026, 2029, 2030, 2126, 2127, 2128, 2308, 2342, 2343, 2346, 2347, 2349, 2350, 2352, 2353, 2355, 2362, 2387, 2388, 2393, 2686, 2758, 2871, 2901, 2976, 3078, 3084, 3170, 3254, 20213, 3202, 579, 1298, 2055, 2118, 2737, 221, 485, 539, 639, 999, 1246, 2022, 2051, 2274, 2691, 2755, 2756, 2757, 2836, 2908, 3252, 3542, 3753, 3820, 3821, 3901, 3904, 3908, 3946, 3950, 20311, 20111, 20176, 20149 & 10.  Octopus 300 S/N: 4242, 3330, 3735 & 1506.  Octopus 600 S/N: 135, 246, 433, 434, 443, 708, 709, 710, 955, 956, 957, 958 & Demo.  Octopus 900 S/N: 2968, 1784, 2023, 2140, 2237, 2883, 2884, 2886, 2967, 3010, 3286, 3753, 1784, 1286, 1937, 2763, 2778, 3746, 3139, 2813, 1143, 1239, 1287, 2233, 2373, 2406, 2566, 2567, 2780, 2889, 2926, 3200, 2890, 2024, 1582, 1619, 1647, 2174, 2586, 2624, 3006, 3009, 3011 & 3372.
510(k) numbers
["K142423"]
PMA numbers
not on file
Product code
NFJ
Firm FEI
1000136533
Firm city
Mason
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register