Stanley Security Solutions Inc: Arial Wireless Water-Resistant Call Pendant The Arial Pendant tag is part of the Arial wireless emergency call management system. It enables residents in assisted living, skilled nursing or independent living to call staff with the press of a button.

Recall number
Z-1499-2017
Recalling firm
Stanley Security Solutions Inc
Product
Arial Wireless Water-Resistant Call Pendant The Arial Pendant tag is part of the Arial wireless emergency call management system. It enables residents in assisted living, skilled nursing or independent living to call staff with the press of a button.
Status
Terminated
Initiated
2016-06-22
Posted
not on file
Terminated
2017-11-20
Reason
Devices were incorrectly programmed during manufacturing therefore depressing the pendant button may result in an alarm not sounding as intended.
Root cause
Software design (manufacturing process)
Action
Stanley Healthcare sent an Important Recall Notice dated June 22, 2016, to all affected customers. The notice instructed customers to discontinue use of the device and to contact the recalling firm to arrange for return of the device. The recalling firm also asked the customer to complete a response form indicating they read, understood and complied with the recall notice. Customers with questions were instructed to contact Arial Customer Service at 888-716-0077. For questions regarding this recall call 781-216-9035.
Quantity
500 individual pendants
Distribution
Nationwide Distribution to AR, AZ, CA, FL, GA, IA, ID, IN, KS, MD, MN, MO, NC, NE, NJ, NY, OH, OR, PA, SD, UT, VA, and WI.
Lot, serial or model codes
Serial numbers WP149149 through WP149398 and WP149649 through WP149898.
510(k) numbers
not on file
PMA numbers
not on file
Product code
ILQ
Firm FEI
1929691
Firm city
Lincoln
Firm state
NE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register