Skip to the content

FDA Medical Device Recalls (openFDA)

Optovue, Inc.: iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA only) and 2016.1.0.127 (out of US) Opthalmic: iVue is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber the eye.

Recall number
Z-1498-2016
Recalling firm
Optovue, Inc.
Product
iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA only) and 2016.1.0.127 (out of US) Opthalmic: iVue is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber the eye.
Status
Terminated
Initiated
2016-03-16
Posted
not on file
Terminated
2017-04-04
Reason
A software anomaly was detected that affects Optovue iVue and iVue 500 with Normative Database that may display images of the same eye rather than right and left if user scrolls on the map.
Root cause
Under Investigation by firm
Action
Recall Letters were sent to customers on March 22, 2016 to inform them of the recall.
Quantity
40
Distribution
PA, CA, IL, HI. MO, CO, GA, ME, IN, AZ, WV, VA, FL, AR, UT, NE, IN, NV, LA, TN, KS, OK. Foreign distribution forthcoming
Lot, serial or model codes
US and foreign (not yet provided) Software version 2016.0.0.127: Serial numbers:  21908, 26507, 26496, 26558, 22738, 26139, 26442, 25720, 25079, 25142, 23298, 21990, 23461, 24199, 25843, 26572, 26550, 22971, 26495, 22652, 26550, 22971, 26495, 22652, 25389, 23309, 23760, 21732-1, 23527, 20923, 21600, 21414, 20271, 23153, 22875, 23211. Software version 2016.1.0.127; Serial numbers forthcoming.
510(k) numbers
["K121739","K133892"]
PMA numbers
not on file
Product code
HLI
Firm FEI
3005950902
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register