FDA Medical Device Recalls (openFDA)
Stryker Instruments Div. of Stryker Corporation: Stryker Instruments 0400-820-000S7 T4 XXL Toga w/T3 Cut, Sterile, O; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.
- Recall number
- Z-1498-2013
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Product
- Stryker Instruments 0400-820-000S7 T4 XXL Toga w/T3 Cut, Sterile, O; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.
- Status
- Terminated
- Initiated
- 2013-05-23
- Posted
- 2013-06-06
- Terminated
- 2014-04-14
- Reason
- Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr
- Root cause
- Nonconforming Material/Component
- Action
- Stryker sent an Urgent Urgent Recall noitifications letter dated May 7, 2013 via certified US mail and E-mail l to all affected customers, distributors and direct accounts. The letter identified the affected products, problem, risk to health, and actions to be taken. Customers were instructed to check inventory, immediately quarantine any affected product, and open each package before discarding to prevent use. Customers were directed to fill out and return a Business Reply Form to Stericycle, Phone: 888-943-2396 Fax: 888-943-4192, Email: stryker5893@stericycle.com verifying directions and confirming recalled product was destroyed. For questions call 269-389-4354 or 269-389-4518.
- Quantity
- 3,980 units ( 398 boxes)
- Distribution
- USA Nationwide Distribution
- Lot, serial or model codes
- 1005098, 1010228, 1012337, 1105367
- 510(k) numbers
- ["K070078"]
- PMA numbers
- not on file
- Product code
- FYA
- Firm FEI
- 1811755
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28