Skip to the content

FDA Medical Device Recalls (openFDA)

Ion Beam Applications S.A.: IBA Proton Therapy System - PROTEUS 235

Recall number
Z-1497-2026
Recalling firm
Ion Beam Applications S.A.
Product
IBA Proton Therapy System - PROTEUS 235
Status
Open, Classified
Initiated
2026-02-03
Posted
2026-03-06
Terminated
not on file
Reason
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Root cause
Software design
Action
On February 3, 2026 Urgent Medical Device Correction letters were sent to customers. Letter provides awareness of the Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. On-site modification of the devices will be performed by IBA with a TSS Configuration update. The update will be deployed at the impacted sites by June 2026.
Quantity
4 units (1 US, 3 OUS)
Distribution
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
Lot, serial or model codes
PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)
510(k) numbers
["K163500"]
PMA numbers
not on file
Product code
LHN
Firm FEI
3000256071
Firm city
Ottignies-Louvain-La-Neuve
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register